Omron BP785 Instruction Manual - Page 45

Error Indicators and Troubleshooting Tips, SYMBOL, CAUSE, CORRECTION - replacement cuffs

Page 45 highlights

ERROR INDICATORS AND TROUBLESHOOTING TIPS ERROR INDICATORS SYMBOL CAUSE Air plug disconnected. CORRECTION Insert the plug securely. Refer to "Applying the Arm Cuff". Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to "Applying the Arm Cuff". Air is leaking from the arm cuff. Replace the cuff with the new one. Cuff over inflated. Repeat measurement. Remain still and do not talk during measurement. Refer to "Taking a Measurement". If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to "Taking a Measurement". The arm cuff was inflated above 299 Do not inflate the cuff above 299 mmHg. Refer to mmHg when inflating the cuff manually. "Taking a Measurement". Movement during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to "Taking a Measurement". Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to "Applying the Arm Cuff". Device error. Contact Customer Service. Sensor may be out of alignment. Contact Customer Service for repair information. Cuff is not applied correctly. Refer to "Applying the Arm Cuff". 45

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50

Contact Customer Service for repair information.
Sensor may be out of alignment.
Refer to “Applying
the Arm Cuff”.
Cuff is not applied correctly.
Air plug disconnected.
Insert the plug securely. Refer to “Applying the
Arm Cuff”.
Arm cuff not applied correctly.
Apply the arm cuff correctly. Refer to “Applying
the Arm Cuff”.
Air is leaking from the arm cuff.
Replace the cuff with the new one.
Cuff over inflated.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
If “E2” appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to “Taking a Measurement”.
The arm cuff was inflated above 299
mmHg when inflating the cuff manually.
Do not inflate the cuff above 299 mmHg. Refer to
“Taking a Measurement”.
Movement during measurement.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering with the arm cuff.
Refer to “Applying the Arm Cuff”.
Device error.
Contact Customer Service.
45
SYMBOL
CAUSE
CORRECTION
ERROR INDICATORS
ERROR INDICATORS AND TROUBLESHOOTING TIPS