Philips HF3331 User manual - Page 25

Guidance And Manufacturer's Declaration - 60 golite blu

Page 25 highlights

26 English GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY:The goLITE BLU is intended for use in the electromagnetic environment specified below.The customer or the user of the goLITE BLU should make sure it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance Electrostatic Discharge (ESD) IEC 61000-4-2 ±(2,4,6) kV contact ±(2,4,6) kV ±(2,4,8) kV air contact ±(2,4,8) kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical Fast Transient/Burst IEC 61000-4-4 ±2 kV for power supply lines ±1 kV for inputoutput lines ±2 kV for power supply lines ±1 kV for input/ output lines Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 ±1 kV differential mode ±2 kV common mode ±1 kV differential mode ±2 kV for common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles 95% dip in UT) for 5 sec 3 A/m 95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles 95% dip in UT) for 5 sec 3 A/m Mains power quality should be that of a typical commercial or hospital environment. If the user of the goLITE BLU requires continued operation during power mains interruptions, it is recommended that the goLITE BLU be powered from an uninterruptible power supply or a battery. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the a.c. mains voltage prior to application of the test level.

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28

26
ENGLISH
GUIDANCE AND MANUFACTURER’S DECLARATION —
ELECTROMAGNETIC IMMUNITY:The goLITE BLU is intended for use in
the electromagnetic environment specified below.The customer or the
user of the goLITE BLU should make sure it is used in such an environment.
I
MMUNITY
T
EST
IEC 60601 T
EST
L
EVEL
C
OMPLIANCE
L
EVEL
E
LECTROMAGNETIC
E
NVIRONMENT
- G
UIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±(2,4,6) kV contact
±(2,4,8) kV air
±(2,4,6) kV
contact
±(2,4,8) kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input-
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV
differential
mode
±2 kV for
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the goLITE BLU requires
continued operation during
power mains interruptions,
it is recommended that the
goLITE BLU be powered from an
uninterruptible power supply or
a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.