C. Calibration and Measurement + T he Celestron Micro CapturePro software includes a measurement feature. To use this feature, you must first perform a measurement calibration. + M ake sure the X,Y Guide is set to "On" by selecting it in the Options Menu. + S et the height of the microscope and focus on your specimen. Then, replace your specimen with the calibration ruler. + R e-adjust the focus and align the calibration ruler so that the lines on the ruler are parallel with the vertical line on the X,Y Guide. + T ake a still image. The image will be displayed in the column on the right side of the screen. + D ouble click the saved image to open it in the Image Preview screen. + Click on the Calibration icon + C lick and drag a line across a "known value" on the ruler image. Try to keep the line as perpendicular to the lines on the ruler as possible. + T he "Adjust magnification" window will pop up as shown here. + In the box "Actual Dimension", enter the "known value" from the calibration ruler 18 I ENGLISH