Calibration And Measurement

Page 5 highlights

4. CALIBRATION AND MEASUREMENT A. CALIBRATION + M ake sure the X,Y Guide is ON (Options/XY Guide). + P lace the included Calibration ruler on the stage under the microscope lens and focus. + T ake a still image making sure the X,Y Guide and the lines on the calibration ruler are parallel. + D ouble-click on the image to enter into measurement interface window. + S elect the Calibration icon and using your cursor, click on a start line on the calibration ruler, hold the mouse button down and drag a dimension line across to the next black line on the calibration ruler. NOTE: T he idea here is to measure a "known" distance on the calibration ruler. For example 1mm or .5 mm. + A dimension line will appear and a dimension will be displayed above the line and a dialogue box will appear. + E nter the "known" distance that you measured, based on the calibration ruler. + T he software will automatically calibrate the magnification rate. + C lick 'OK' and the adjusted magnification rate will show in the upper right corner. NOTE: S ave this magnification rate, so that you may use it for future settings. + N ow you can take measurements on the saved image. ENGLISH | 5

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50

ENGLISH |
5
+
Make sure the X,Y Guide is ON (Options/XY Guide).
+
Place the included Calibration ruler on the stage under the
microscope lens and focus.
+
Take a still image making sure the X,Y Guide and the lines
on the calibration ruler are parallel.
+
Double-click on the image to enter into measurement
interface window.
A. CALIBRATION
4. CALIBRATION AND MEASUREMENT
+
A dimension line will appear and a dimension will be displayed
above the line and a dialogue box will appear.
+
Enter the “known” distance that you measured, based on the
calibration ruler.
+
The software will automatically calibrate the magnification rate.
+
Click ‘OK’ and the adjusted magnification rate will show in the
upper right corner.
NOTE:
Save this magnification rate, so that you may use it for
future settings.
+
Now you can take measurements on the saved image.
+
Select the Calibration icon
and using your cursor,
click on a start line on the calibration ruler, hold the mouse
button down and drag a dimension line across to the next
black line on the calibration ruler.
NOTE:
The idea here is to measure a “known” distance on the
calibration ruler. For example 1mm or .5 mm.