Fluke 117 Fluke 114,115,117 Manual - Page 21
Surge Protection, Temperature Coefficient - specifications
View all Fluke 117 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
True-rms Multimeters General Specifications General Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Extended specifications are available at www.Fluke.com. Maximum voltage between any terminal and earth ground 600 V Surge Protection 6 kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2 W Fuse for A input (115 & 117 only 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) Display Digital: 6,000 counts, updates 4/sec Bar Graph: 33 segments, updates 32/sec Temperature Operating: -10 °C to + 50 °C Storage: -40 °C to + 60 °C Temperature Coefficient 0.1 x (specified accuracy)/°C (28 °C) Operating Altitude 2,000 meters Battery 9 Volt Alkaline, NEDA 1604A / IEC 6LR61 Battery Life Alkaline: 400 hours typical, without backlight Safety Compliances Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1 19