Fluke 1551A-12 Calibration Guide

Fluke 1551A-12 Manual

Fluke 1551A-12 manual content summary:

  • Fluke 1551A-12 | Calibration Guide - Page 1
    1551A Ex/1552A Ex Stik Thermometer Calibration Manual November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Fluke 1551A-12 | Calibration Guide - Page 2
    Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service
  • Fluke 1551A-12 | Calibration Guide - Page 3
    Table of Contents Title Page Introduction...1 How to Contact Fluke 1 Safety Information 2 ATEX Safety Information 3 Symbols ...3 Approvals ...4 Specifications...4 Calibration ...5 Overview ...5 Recommended Equipment 6 Initial Setup 6 Record as-found Calibration Parameters 7 Evaluate as-
  • Fluke 1551A-12 | Calibration Guide - Page 4
    1551A Ex/1552A Ex Calibration Manual ii
  • Fluke 1551A-12 | Calibration Guide - Page 5
    Measurement Errors 8 6. Low-Adjustment Temperature 8 7. Zero Adjustment Temperature 9 8. High Adjustment Temperature 9 9. As-left Measurement Errors 10 10. As-left Calibration Parameters 11 11. Approved Batteries 12 iii
  • Fluke 1551A-12 | Calibration Guide - Page 6
    1551A Ex/1552A Ex Calibration Manual iv
  • Fluke 1551A-12 | Calibration Guide - Page 7
    List of Figures Figure Title Page 1. Battery Replacement 12 v
  • Fluke 1551A-12 | Calibration Guide - Page 8
    1551A Ex/1552A Ex Calibration Manual vi
  • Fluke 1551A-12 | Calibration Guide - Page 9
    same hazardous locations as analog thermometers. For complete usage instructions, see the 1551A Ex/1552A Ex Users Manual. How to Contact Fluke To contact Fluke, call one of the following telephone numbers: • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) • Calibration/Repair USA: 1-888-99
  • Fluke 1551A-12 | Calibration Guide - Page 10
    1551A Ex/1552A Ex Calibration Manual Safety Information A Warning identifies conditions and actions that Ex- hazardous. • Use only specified approved batteries in the Product. Refer to the battery replacement instructions for a list of approved batteries. • Do not use and disable the Product if it
  • Fluke 1551A-12 | Calibration Guide - Page 11
    Fluke Corporation. Refer to "How to Contact Fluke". Substitution of components will impair suitability for hazardous locations. ATEX Safety Information This manual full manual before you use the Product. To make sure of safe Product operation, fully follow all instructions and warnings in the manual.
  • Fluke 1551A-12 | Calibration Guide - Page 12
    1551A Ex/1552A Ex Calibration Manual Approvals The Product meets the above requirements using: EN 60079-0:2006 (IEC 60079-11:2006 (EQV) + CORRIGENDUM DEC. 2006 (EQV)) Manufactured by Fluke Corporation: Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. P.O. Box 1186 5602 BD Eindhoven
  • Fluke 1551A-12 | Calibration Guide - Page 13
    4 in. x 2 in. x 1 in. Probe Size 1551A-9: 4.8 mm x 229 mm (3/16 in. x 9 in.) 1551A-12: 6.35 mm x 305 mm (1/4 in. x 12 in.) 1551A-20: 6.35 mm x 508 mm (1/4 in. x 20 in personnel. The technician should be familiar with the 1551A Ex/1552A Ex Users Manual. The calibration must be performed in the
  • Fluke 1551A-12 | Calibration Guide - Page 14
    1551A Ex/1552A Ex Calibration Manual Recommended Equipment The equipment listed in Table 2 will satisfy the calibration requirements: Table 2. Recommended Equipment Equipment Recommended Model Temperature Baths Fluke 6331 40 °C to 300 °C Hot Bath Fluke 7381 -80 °C to 110 °C Cold Bath Fluke
  • Fluke 1551A-12 | Calibration Guide - Page 15
    it be tested near the five temperatures listed in Table 4, depending on the model. Table 4. Test Temperatures Point 1 2 3 4 5 -50 °C -25 °C 0 °C 100 °C 157 °C 1551A Ex -80 °C -25 °C 0 °C 100 °C 300 °C 1552A Ex At each of the test temperatures, readings are recorded of the temperature of the
  • Fluke 1551A-12 | Calibration Guide - Page 16
    Ex/1552A Ex Calibration Manual At each calibration point the measurement error may be calculated by subtracting the reference thermometer temperature from the 1551A Ex/1552A Ex temperature. The temperatures and measurement errors may be recorded in Table 5. Table 5. As-found Measurement Errors
  • Fluke 1551A-12 | Calibration Guide - Page 17
    of the Product to settle and then verify that the readings match the temperature of the reference thermometer well within the specified accuracy of the 1551A Ex/1552A Ex. Adjustment at the High Calibration Point The final adjustment is made for the high end of the temperature range of the Product
  • Fluke 1551A-12 | Calibration Guide - Page 18
    1551A Ex/1552A Ex Calibration Manual The command will set the value of the CAL: verify that the readings match the temperature of the reference thermometer well within the specified accuracy of the 1551A Ex/1552A Ex. Set the Calibration Date The date of calibration should be stored in the Product
  • Fluke 1551A-12 | Calibration Guide - Page 19
    static awareness A Message From Fluke Corporation Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by: 1. Knowing that there is a problem. 2. Learning the guidelines for
  • Fluke 1551A-12 | Calibration Guide - Page 20
    5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT. 8. WHEN REMOVING PLUG-IN ASSEMBLIES HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES. 6. DO NOT
  • Fluke 1551A-12 | Calibration Guide - Page 21
    Stik Thermometer Maintenance Maintenance Clean the Product WCaution To prevent possible damage to the product or to equipment under test, do not use abrasive cleaners. They will damage the case. To clean the Product, use a cloth with a mild cleaning solution. Battery Replacement XWWarning To prevent
  • Fluke 1551A-12 | Calibration Guide - Page 22
    1551A Ex/1552A Ex Calibration Manual To change the batteries, see Figure 1: 1. Use a flat-blade screwdriver to remove the battery door screw. 2. Remove the Table 11. Approved Batteries Manufacturer Duracell Rayovac Energizer Panasonic Type LR03/MN2400 LR03/824 LR03/E92 LR03X gjo003.eps 12
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22

November 2010
© 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
1551A Ex/1552A Ex
Stik Thermometer
Calibration Manual