Fluke 1551A-12 Calibration Guide - Page 12

Approvals, Specifications

Page 12 highlights

1551A Ex/1552A Ex Calibration Manual Approvals The Product meets the above requirements using: EN 60079-0:2006 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES -- PART 0: GENERAL REQUIREMENTS (IEC 60079-0:2004 (MOD)) EN 60079-11:2007 EXPLOSIVE ATMOSPHERES -- PART 11: EQUIPMENT PROTECTION BY INTRINSIC SAFETY "I" (IEC 60079-11:2006 (EQV) + CORRIGENDUM DEC. 2006 (EQV)) Manufactured by Fluke Corporation: Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. P.O. Box 1186 5602 BD Eindhoven The Netherlands Specifications (Ambient: 23 °C ±5 °C) Measurement Range 1551A Ex 50 °C to 160 °C (−58 °F to 320 °F) 1552A Ex 80 °C to 300 °C (−112 °F to 572 °F) Accuracy (1 year 0.05 °C (0.09 °F) Resolution Selectable (0.1, 0.01, 0.001) factory default is 0.01 Sample Rate User selectable 0.5/sec, 1/sec or 2/sec factory default is 1/sec Operating Temperature Range of Readout 10 °C to 50 °C (14 °F to 122 °F) Probe Response Time Approximately 20 seconds EMC Compliance EN61326:2006 Annex C CISPR 11, Edition 5.0-2009 Class "B" Humidity Range 0 to 95 % RH Non-condensing Storage Temperature Range 20 °C to 60 °C (−4 °F to 140 °F) Enclosure Rating IP50 Readout Temperature Coefficient Add ±10 ppm/°C of full scale temperature from −10 °C to 18 °C and 28 °C to 50 °C Probe Temperature Coefficient 0.00385 Ω/Ω/°C nominal 4

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22

1551A Ex/1552A Ex
Calibration Manual
4
Approvals
The Product meets the above requirements using:
EN 60079-0:2006 ELECTRICAL APPARATUS FOR
EXPLOSIVE GAS ATMOSPHERES -- PART
0: GENERAL REQUIREMENTS (IEC
60079-0:2004 (MOD))
EN 60079-11:2007
EXPLOSIVE ATMOSPHERES -- PART 11:
EQUIPMENT PROTECTION BY INTRINSIC
SAFETY "I" (IEC 60079-11:2006 (EQV) +
CORRIGENDUM DEC. 2006 (EQV))
Manufactured by Fluke Corporation:
Fluke Corporation
P.O. Box 9090
Everett, WA
98206-9090
U.S.A.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Specifications
(Ambient: 23
°
C
±
5
°
C)
Measurement Range
1551A Ex
.........................................................................
50
°
C to 160
°
C (
58
°
F to 320
°
F)
1552A Ex
.........................................................................
80
°
C to 300
°
C (
112
°
F to 572
°
F)
Accuracy (1 year)
.....................................................................
±
0.05
°
C (0.09
°
F)
Resolution
.................................................................................
Selectable (0.1, 0.01, 0.001) factory default is 0.01
Sample Rate
.............................................................................
User selectable 0.5/sec, 1/sec or 2/sec factory default is
1/sec
Operating Temperature Range of Readout:
..........................
10
°
C to 50
°
C (14
°
F to 122
°
F)
Probe Response Time
..............................................................
Approximately 20 seconds
EMC Compliance
.......................................................................
EN61326:2006 Annex C
CISPR 11, Edition 5.0-2009
Class “B”
Humidity Range
.........................................................................
0 to 95 % RH Non-condensing
Storage Temperature Range
....................................................
20
°
C to 60
°
C (
4
°
F to 140
°
F)
Enclosure Rating
......................................................................
IP50
Readout Temperature Coefficient
...........................................
Add
±
10 ppm/
°
C of full scale temperature from
10
°
C to
18
°
C and 28
°
C to 50
°
C
Probe Temperature Coefficient
..............................................
0.00385
Ω
/
Ω
/
°
C nominal