Fluke 190-104/AM Product Manual - Page 8

Safety

Page 8 highlights

Accuracy Input Range Intrinsic Accuracy Aux Adapter + Instrument (% of Reading + % of Range) Direct Input ±10 V see instrument specification Voltage divider input ±1000 V ±(0.5% + 0.002%) Note: Environmental Reference Conditions: 23 °C ±5 °C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10

Accuracy
Temperature Coefficient
Direct Input
........................
See specification for Logger
Voltage Divider Input
.........
150 ppm/°C (for temperatures <18 °C and
>28 °C)
Safety
Direct Input
Maximum Voltage between
Direct Input Terminals and
Earth Ground
..................
30 V
Voltage Divider Input
Maximum Voltage between
4 mm Safety Socket and
Earth Ground
..................
1000 V
IEC61010-1
.......................
Pollution Degree 2
IEC61010-2-030
................
CAT III 1000 V
Environmental
IP Rating
.........................
IEC 60529:IP50
Temperature
Operating
....................
-10 °C to +50 °C (14 °F to 122 °F)
Storage
.......................
-20 °C to +60 °C (-4 °F to +140 °F)
Operating Humidity
.........
<10 °C (<50 °F) non condensing
10 °C to 30 °C (50 °F to 86 °F)
95 %
30 °C to 40 °C (86 °F to 104 °F)
75 %
40 °C to 50 °C (104 °F to 122 °F)
45 %
Altitude
Operating
......................
2000 m (>2000 m to
4000 m, derate to
CAT IV 300 V / CAT III 600 V / CAT II
1000 V)
Storage
.......................
12 000 m
Input
Range
Intrinsic Accuracy Aux
Adapter + Instrument
(% of Reading + % of
Range)
Direct Input
±10 V
see instrument specification
Voltage divider input
±1000 V
±(0.5% + 0.002%)
Note:
Environmental Reference Conditions: 23 °C ±5 °C, instrument
operating for at least 30 minutes, no external electrical/magnetic
field, RH <65 %.