Fluke 572 FE 572 Users Manual - Page 22
Specifications, FACTORY DEFAULTS
View all Fluke 572 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 22 highlights
572 Specifications Temperature range Display Resolution Accuracy Ambient derating Repeatability Response Time (95%) Hot Spot Detection (30%) Spectral Range Optical Resolution (Standard Focus) Optical Resolution (Close Focus) Ambient Operating Range Storage Temperature (without batteries) Power Battery Life Dimensions Tripod Mount -30 to 900°C (-25 to 1600°F) 0.1°C (0.2°F) ± 0.75% of reading or ± 1K (± 1.5°F), whichever is greater at 25°C (77°F) ambient temperature, ± 2°K (± 4°F) for targets below -5°C (23°F) < 0.05K/K or < 0.05 %/K, whichever is greater at +25°C (77°F) ± 25K (± 45°F) ±0.5% of reading or ±1°C (±2°F), whichever is greater 250 mSec 85 mSec 8 to 14 µm 60 : 1 (19mm spot size at 1.15M.) (0.75 in. spot size at 3.8 feet) 50 : 1 (6mm spot size at 0.3M.) (0.24 in. spot size at 0.98 feet) 0 to 50°C (32 to 122°F) -20 to 50°C (-4 to 122°F) 2 x 1.5 V Alkaline Type AA 13 hrs. (50% laser and 50% backlight used) 200 x 170 x 50 mm (7.9 x 6.7 x 2 inches) 1/4"-20 UNC Factory defaults Default FACTORY DEFAULTS Range Emissivity/Gain Hi Alarm 0.95 50°C (122°F) 0.10 to 1.50, in steps of 0.01 -30 to 900°C (-25 to 1600°F) 22