Fluke 700G07 Product Manual - Page 21
Specifications
View all Fluke 700G07 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
Pressure Gauge Specifications Specifications Available Input Ranges See Ranges and Resolution table for available ranges in psi plus equivalent ranges and resolution for all engineering units. Accuracy 700G Ranges Positive Pressure 0.05 % FS Positive Pressure (700G01, 700G02 0.1 % FS Vacuum 0.1 % FS Temperature Compensation 15 °C to 35 °C (59 °F to 95 °F) to rated accuracy Note: For temperatures from -10 °C to 15 °C and 35 °C to 55 °C, add .003 % FS/°C Accuracy 700RG Ranges Positive Pressure 0.04 % rdg ±0.01 % FS Vacuum (700RG05 0.05 % FS Vacuum (all other ranges 0.1 % FS Temperature Compensation 0 °C to 50 °C (32 °F to 122 °F) to rated accuracy Note For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add .005 % FS/°C. 13 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664