Fluke 789 Calibration Manual - Page 12
Specifications, DC Volts Measurement
View all Fluke 789 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 12 highlights
789 Calibration Manual Specifications All specifications apply from +18 °C to +28 °C unless stated otherwise. All specifications assume a 5-minute warm-up period. The standard specification interval is 1 year. Note "Counts" refers to the number of increments or decrements of the least significant digit. DC Volts Measurement Range (V dc) 4.000 40.00 400.0 1000 Resolution 0.001 V 0.01 V 0.1 V 1 V Accuracy, ±(% of Reading + Counts) 0.1 % + 1 0.1 % + 1 0.1 % + 1 0.1 % + 1 Input impedance: 10 MΩ (nominal), 60 dB at 50 Hz or 60 Hz Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz Overvoltage protection: 1000 V DC Millivolts Measurement Range (mV dc) 400.0 Resolution 0.1 mV Accuracy, ±(% of Reading + Counts) 0.1 % + 2 AC Volts Measurement Range (ac) Resolution Accuracy, ±(% of Reading + Counts) 50 Hz to 60 Hz 45 Hz to 200 Hz 200 Hz to 500 Hz 400.0 mV 0.1 mV 0.7 % + 4 1.2 % + 4 7.0 % + 4 4.000 V 40.00 V 400.0 V 1000 V 0.001 V 0.01 V 0.1 V 1 V 0.7 % + 2 0.7 % + 2 0.7 % + 2 0.7 % + 2 1.2 % + 4 1.2 % + 4 1.2 % + 4 1.2 % + 4 7.0 % + 4 7.0 % + 4 7.0 % + 4 7.0 % + 4 Specifications are valid from 5 % to 100 % of amplitude range. AC conversion: true rms Maximum crest factor: 3 (between 50 and 60 Hz) For non-sinusoidal waveforms, add ±(2 % reading + 2 % f.s.) typical Input impedance: 10 MΩ (nominal), 60 dB at dc, 50 Hz, or 60 Hz 4