Fluke 789 Calibration Manual - Page 22

Current Measurement, terminals on the UUT.

Page 22 highlights

789 Calibration Manual Current Measurement 1. Put the calibrator in Standby (STBY) mode. W 2. Put the UUT rotary switch in the position. 3. Connect the calibrator to the COM and mA terminals on the UUT as shown in Figure 4. 4. Apply the values from the calibrator shown in Table 5 and compare the readings on the UUT to the acceptable readings shown. 5. Connect the calibrator to the COM and A terminals on the UUT. 6. Apply the values from the calibrator shown in Table 5 and Table 6 and compare the readings on the UUT to the acceptable readings shown. 7. Press J (BLUE) on the UUT to toggle to ac amps. 8. Apply the values from the calibrator shown in Table 7 and compare the readings on the UUT to the acceptable readings shown. UUT 5500A Figure 4. Current Measurement Test Connections adm003F.EPS 14

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

789
Calibration Manual
14
Current Measurement
1.
Put the calibrator in Standby (
STBY
) mode.
2.
Put the UUT rotary switch in the
W
position
.
3.
Connect the calibrator to the
COM
and
mA
terminals on the UUT as shown in
Figure 4.
4.
Apply the values from the calibrator shown in Table 5 and compare the readings on
the UUT to the acceptable readings shown.
5.
Connect the calibrator to the
COM
and
A
terminals on the UUT.
6.
Apply the values from the calibrator shown in Table 5 and Table 6 and compare the
readings on the UUT to the acceptable readings shown.
7.
Press
J
(BLUE) on the UUT to toggle to ac amps.
8.
Apply the values from the calibrator shown in Table 7 and compare the readings on
the UUT to the acceptable readings shown.
UUT
5500A
adm003F.EPS
Figure 4. Current Measurement Test Connections