    ® 902 HVAC Clamp Meter Users Manual PN 2547887 May 2006 Rev. 1, 3/07 © 2006-2007 Fluke Corporation. All rights reserved. Printed in China. All product names are trademarks of their respective companies.
  • Fluke 902 | FE 902 Users Manual - Page 2
    to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem. THIS WARRANTY IS YOUR ONLY
  • Fluke 902 | FE 902 Users Manual - Page 3
    Contents Title Page Introduction 1 Contacting Fluke 2 Safety Information 3 Symbols 5 Getting Acquainted with the Meter 6 Using the Meter 10 AC and DC Voltage Measurement 10 Resistance and Continuity 11 Microamps µA Measurement 12 Temperature 13 Capacitance 16 AC Current Measurement 16
  • Fluke 902 | FE 902 Users Manual - Page 4
    902 Users Manual ii
  • Fluke 902 | FE 902 Users Manual - Page 5
    List of Tables Table Title Page 1. 902 HVAC Clamp Meter Features 7 2. Display Features 9 List of Figures Figure Title Page 1. 902 HVAC Clamp Meter Features 6 2. Display Features 8 3. Testing a Flame Rod 13 4. Temperature Measurement 15 5. Proper AC Current Measurement 17 6. Battery
  • Fluke 902 | FE 902 Users Manual - Page 6
    902 Users Manual iv
  • Fluke 902 | FE 902 Users Manual - Page 7
    rod testing) • AC and DC voltages • Capacitance • Resistance • Continuity • Temperature in both Celsius (°C) and Fahrenheit (°F) The Meter comes with: • Two AA alkaline batteries (installed) • Users Manual • Soft carrying case • TL75 Test Leads (one pair) • 80BK Integrated DMM Temperature Probe 1
  • Fluke 902 | FE 902 Users Manual - Page 8
    902 Users Manual Contacting Fluke To contact Fluke, call one of the following telephone numbers: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or visit Fluke's
  • Fluke 902 | FE 902 Users Manual - Page 9
    damage the Meter or the equipment under test. XWRead First: Safety Information To ensure safe operation and service of the Meter, follow these instructions: • Read the Users Manual before use and follow all safety instructions. • Use the Meter only as specified in the Users Manual; otherwise, the
  • Fluke 902 | FE 902 Users Manual - Page 10
    902 Users Manual • Use caution when working with voltages above 60 V dc or 30 V ac rms or 42 V ac peak. Such voltages pose a shock hazard. • Clean the case with a damp cloth and mild detergent only. Do not use abrasives or solvents. • To avoid false readings that can lead to electrical shock and
  • Fluke 902 | FE 902 Users Manual - Page 11
    HVAC Clamp Meter Symbols Symbols The following symbols are found on the Meter or in this manual. , May be used on hazardous live conductors W Risk of danger. Important information. See Users Manual this product as unsorted municipal waste. Contact Fluke or a qualified recycler for disposal.
  • Fluke 902 | FE 902 Users Manual - Page 12
    902 Users Manual Getting Acquainted with the Meter Refer to Figures 1 and 2 and Tables 1 and 2 to become more acquainted with the Meter's features. COM CAT µAV 600V efu0001.eps Figure 1. 902 HVAC Clamp Meter Features 6
  • Fluke 902 | FE 902 Users Manual - Page 13
    with the Meter Table 1. 902 HVAC Clamp Meter Features Number Description A Backlight Button B Jaw Release C Tactile Barrier D Jaws E Hold Button F Rotary Switch: V DC and AC voltage P Resistance and continuity G DC microamps F L K Degrees Fahrenheit / degrees Celsius Capacitance AC
  • Fluke 902 | FE 902 Users Manual - Page 14
    902 Users Manual C D EF G B A Figure 2. Display Features efu0006.eps 8
  • Fluke 902 | FE 902 Users Manual - Page 15
    HVAC Clamp Meter Getting Acquainted with the Meter Number Table 2. Display Features Indication Battery indicator -The batteries are low and need to be changed. XWWarning: To avoid false A readings, which could lead °C - Degrees Celsius e - Ohms G M - Microamps N - Microfarads DC - Direct
  • Fluke 902 | FE 902 Users Manual - Page 16
    902 Users Manual Using the Meter AC and DC Voltage Measurement To measure AC or DC voltage: 1. Insert the test leads into the Meter. V 2. Turn the rotary switch to . 3. Press A to choose AC or DC voltage. The display reflects the chosen voltage mode. 4. Use the test leads to take the measurement.
  • Fluke 902 | FE 902 Users Manual - Page 17
    HVAC Clamp Meter Using the Meter Resistance and Continuity To measure resistance or continuity: XW Warning To avoid false readings that can lead to electrical shock and injury, de-energize the circuit before taking the measurement. 1. Insert the test leads into the Meter. P 2. Turn the rotary switch
  • Fluke 902 | FE 902 Users Manual - Page 18
    902 Users Manual Microamps µA Measurement The µA dc (G) function on the Meter is primarily for HVAC flame rod testing. To test a heating system 3. Turn the rotary switch on the Meter to G. 4. Using alligator clips, connect test leads between the flame sensor probe and control-module wire. 5. Turn
  • Fluke 902 | FE 902 Users Manual - Page 19
    HVAC Clamp Meter Using the Meter Burner Figure 3. Testing a Flame Rod efu0004.eps Temperature The Meter measures temperature in either Celsius (°C) or Fahrenheit (°F). To measure temperature (refer to Figure 4): 13
  • Fluke 902 | FE 902 Users Manual - Page 20
    902 Users Manual 1. Connect the 80BK Integrated DMM Temperature Probe to the input jacks noting correct polarity of the probe. 2. Turn the rotary switch to F. 3. Press A to select °C or °F. The display reflects the chosen temperature mode. 4. Position the probe to take the measurement. The reading
  • Fluke 902 | FE 902 Users Manual - Page 21
    HVAC Clamp Meter Using the Meter Figure 4. Temperature Measurement efu0005.eps 15
  • Fluke 902 | FE 902 Users Manual - Page 22
    Meter's rotary switch to capacitance (L). If the capacitor requires more discharging, diSC is displayed while the capacitor discharges. When measuring, be sure to note the correct polarity of the capacitor. AC Current Measurement XW Warning To avoid electrical shock and injury: • Remove Test Leads
  • Fluke 902 | FE 902 Users Manual - Page 23
    HVAC Clamp Meter Using the Meter efu0003.eps Figure 5. Proper AC Current Measurement 17
  • Fluke 902 | FE 902 Users Manual - Page 24
    902 Users Manual Backlight Press C to toggle the backlight on and off. The backlight automatically turns off after 2 minutes. To disable the automatic 2-minute backlight timeout, hold down Cwhile turning the Meter is detected, the Meter beeps. To use this feature: 1. Put the Meter into the desired
  • Fluke 902 | FE 902 Users Manual - Page 25
    HVAC Clamp Meter Using the Meter Display HOLD XWWarning To avoid possible electric shock or personal injury, when Display HOLD is activated, be aware that the display will not change when you apply a different voltage. In the Display HOLD mode, the Meter freezes the display. The Meter also beeps
  • Fluke 902 | FE 902 Users Manual - Page 26
    shock or personal injury, repairs or servicing not covered in this manual should be performed only by qualified personnel. Cleaning the Meter XWWarning To avoid electrical shock, remove any input signals before cleaning. W Caution To avoid damaging the Meter, do not use aromatic hydrocarbons or
  • Fluke 902 | FE 902 Users Manual - Page 27
    HVAC Clamp Meter Maintenance Battery Replacement XWWarning To avoid false readings that could lead to possible electric shock or personal injury, replace the batteries as soon as the low battery indicator (B) appears. Disconnect the test leads before replacing the batteries. To replace the batteries
  • Fluke 902 | FE 902 Users Manual - Page 28
    902 Users Manual Figure 6. Battery Replacement efu0002.eps 22
  • Fluke 902 | FE 902 Users Manual - Page 29
    HVAC Clamp Meter Specifications Specifications Electrical Specifications Function Range Resolution Accuracy Voltage DC 0 - 600 V 0.1 V 1 % ± 5 counts Voltage AC (True Rms) 0 - 600 V 0.1 V 1 % ± 5 counts (50/60 Hz) Current AC (True Rms) 0 - 600 A 0.1 A 2.0 % ± 5 counts (50/60 Hz)
  • Fluke 902 | FE 902 Users Manual - Page 30
    902 Users Manual General Specifications Operating Temperature -10 °C to +50 °C Storage Temperature -40 °C to +60 °C Operating Humidity Non condensing (< 10 °C) 90 % RH (10 °C to 30 °C) 75 % RH (30 °C to 40 °C) 45 % RH (40 °C to 50 °C) (Without Condensation) Operating Altitude 2500 meters
  • Fluke 902 | FE 902 Users Manual - Page 31
    Size (H X W X L) Weight Design Standards and Compliance Agency Approvals Over-voltage Category HVAC Clamp Meter Specifications 9.1 x 3.8 x 1.7 inches (240 x 80 x 40 mm) 1.1 lb (310 g) IEC 61010, IEC 61010-2032,CE P ) ; s N10140 600 V, CAT III per IEC 1010-1 Power Requirements CAT III equipment
  • Fluke 902 | FE 902 Users Manual - Page 32
    902 Users Manual 26
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

®