Fluke 9143-A-P-156 Product Manual - Page 26

Block Temperature Indicator 10 [Patent Pending], Heat Source Temperature 1, Set-point Temperature 2

Page 26 highlights

914X Field Metrology Wells Parts and Controls Block Temperature Indicator (10) [Patent Pending] The Block Temperature Indicator lamp allows users to know when the block temperature is safe (50°C to 60°C) to remove inserts or move the Field Metrology Well. The indicator light is lit continuously once the block has exceeded approximately 50°C (varies 50°C to 60°C). The indicator light stays lit until the block cools to less than approximately 50°C. If the instrument is disconnected from mains power, the indicator light flashes until the block temperature is less than approximately 50°C. Figure 3 Display panel and keys 3.2.2 Display The front panel display is shown in detail in Figure 4 on opposite page. Heat Source Temperature (1) The most recent block temperature measurement is shown in large digits in the box at the top of the screen. Set-point Temperature (2) The current set-point temperature is displayed just below the Process Temperature. Reference Thermometer Temperature (3) [-P models only] When installed, the most recent reference thermometer measurement is shown on the screen. 20 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42

914X Field Metrology Wells
Parts and Controls
20
Block Temperature Indicator (10) [Patent Pending]
The Block Temperature Indicator lamp allows users to know when the block temperature is
safe (50°C to 60°C) to remove inserts or move the Field Metrology Well. The indicator light is
lit continuously once the block has exceeded approximately 50°C (varies 50°C to 60°C). The
indicator light stays lit until the block cools to less than approximately 50°C. If the instrument
is disconnected from mains power, the indicator light flashes until the block temperature is less
than approximately 50°C.
Display panel and keys
Figure 3
Display
3.2.2
The front panel display is shown in detail in Figure 4 on opposite page.
Heat Source Temperature (1)
The most recent block temperature measurement is shown in large digits in the box at the top of
the screen.
Set-point Temperature (2)
The current set-point temperature is displayed just below the Process Temperature.
Reference Thermometer Temperature (3) [-P models only]
When installed, the most recent reference thermometer measurement is shown on the screen.