Fluke P5514B-2700G-1 Product Manual - Page 6
Specifications, Operating Fluid Compatibility
View all Fluke P5514B-2700G-1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 6 highlights
P5514B-70M/P5514B-70M-EP Users Manual Specifications Pressure Range Pressure Connections Adapters Provided P5514B-70M Seal Material P5514B-70M-EP Seal Material Screw Press Displacement Reservoir Volume Fine Adjustment Displacement Size (H,W,D) Weight Safety 70 MPa (10 000 psi) M20 1/4 NPT Female (quantity 2), 1/2 NPT Female, 1/4 BSP Female, 1/2 BSP Female Viton® Ethylene Propylene 20 cm3(1.2 in3) 75 cm3 (4.5 in3) 0.06 cm3 (0.004 in3) 150 x 330 x 280 mm (6 x 3 x 11 in) 5 kg (11 lb) 2014/68/EU, Pressure Equipment Directive (PED) - Article 4, Paragraph 3 Operating Fluid Compatibility The standard system uses Viton O-ring seals and is for use with a wide range of fluids. Compatible fluids include, but are not limited to, mineral oils, water, and solvents. For compatibility with phosphate esters, a model with ethylene propylene seals is available. See Table 1. Table 1. Operating Fluid Compatibility Model P5514B-70M P5514B-70M-EP O Rings Viton Ethylene Propylene 1. Rotate the reservoir dust cover 1/2 turn. 2. Fill reservoir approximately 3/4 full with the appropriate fluid. 3. Rotate dust cover back to cover reservoir. WCaution To avoid damage to the instrument, check the quality of the operating fluid during use. If the fluid becomes discolored, cloudy or particles appear in the reservoir, drain and flush the system with clean fluid. 2 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664