Fluke P5514B-2700G-1 Product Manual - Page 6

Specifications, Operating Fluid Compatibility

Page 6 highlights

P5514B-70M/P5514B-70M-EP Users Manual Specifications Pressure Range Pressure Connections Adapters Provided P5514B-70M Seal Material P5514B-70M-EP Seal Material Screw Press Displacement Reservoir Volume Fine Adjustment Displacement Size (H,W,D) Weight Safety 70 MPa (10 000 psi) M20 1/4 NPT Female (quantity 2), 1/2 NPT Female, 1/4 BSP Female, 1/2 BSP Female Viton® Ethylene Propylene 20 cm3(1.2 in3) 75 cm3 (4.5 in3) 0.06 cm3 (0.004 in3) 150 x 330 x 280 mm (6 x 3 x 11 in) 5 kg (11 lb) 2014/68/EU, Pressure Equipment Directive (PED) - Article 4, Paragraph 3 Operating Fluid Compatibility The standard system uses Viton O-ring seals and is for use with a wide range of fluids. Compatible fluids include, but are not limited to, mineral oils, water, and solvents. For compatibility with phosphate esters, a model with ethylene propylene seals is available. See Table 1. Table 1. Operating Fluid Compatibility Model P5514B-70M P5514B-70M-EP O Rings Viton Ethylene Propylene 1. Rotate the reservoir dust cover 1/2 turn. 2. Fill reservoir approximately 3/4 full with the appropriate fluid. 3. Rotate dust cover back to cover reservoir. WCaution To avoid damage to the instrument, check the quality of the operating fluid during use. If the fluid becomes discolored, cloudy or particles appear in the reservoir, drain and flush the system with clean fluid. 2 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10

P5514B-70M/P5514B-70M-EP
Users Manual
2
Specifications
Operating Fluid Compatibility
The standard system uses Viton O-ring seals and is for use with a wide range of fluids.
Compatible fluids include, but are not limited to, mineral oils, water, and solvents. For
compatibility with phosphate esters, a model with ethylene propylene seals is available. See
Table
1
.
Table 1. Operating Fluid Compatibility
1.
Rotate the reservoir dust cover 1/2 turn.
2.
Fill reservoir approximately 3/4 full with the appropriate fluid.
3.
Rotate dust cover back to cover reservoir.
W
Caution
To avoid damage to the instrument, check the quality of the operating fluid during
use. If the fluid becomes discolored, cloudy or particles appear in the reservoir,
drain and flush the system with clean fluid.
Pressure Range
70 MPa (10 000 psi)
Pressure Connections
M20
Adapters Provided
1/4 NPT Female (quantity 2), 1/2 NPT Female, 1/4 BSP Female, 1/2 BSP
Female
P5514B-70M Seal Material
VitonĀ®
P5514B-70M-EP Seal
Material
Ethylene Propylene
Screw Press
Displacement
20 cm
3
(1.2 in
3
)
Reservoir Volume
75 cm
3
(4.5 in
3
)
Fine Adjustment
Displacement
0.06 cm
3
(0.004 in
3
)
Size (H,W,D)
150 x 330 x 280 mm (6 x 3 x 11 in)
Weight
5 kg (11 lb)
Safety
2014/68/EU, Pressure Equipment Directive (PED) - Article 4, Paragraph 3
Model
O Rings
P5514B-70M
Viton
P5514B-70M-EP
Ethylene Propylene
Shop for Fluke products online at: