SECTION 4-TYPICAL PRODUCT PARAMETERS Change in maximum recommended flow when back pressure of 7kPa is applied: 0.7 L/min Filters: Cabinet, Outlet HEPA and Compressor Inlet Safety System: Width: Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System (SensO2 Model) Low Flow Alarm. 15 in ± 3/8 in (38.1 cm ± 1 cm) Height: 23 in ± 3/8 in (58.4 cm ± 1 cm) Depth: 12 in ± 3/8 in (30.5 cm ± 1 cm) Weight: Shipping Weight: Operating Ambient Temperature: 43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto2 Models 45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto2W Models 40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto2V Model 48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto2 Models 50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto2W Models 45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto2V Model 50°F - 95°F (10°C - 35°C) at 20-60% relative humidity Perfecto2 Models and Perfecto2V Model Exhaust: Oxygen Output Temperature: Cabinet: Regulatory Listing: Electrical: 50°F - 90°F (10°C - 32°C) at 20-60% relative humidity Perfecto2W Models Less than Ambient + 35°F (+19°C) Less than Ambient + 6°F (+3°C) Impact Resistant flame-retardant plastic cabinet that conforms to UL94V-0. Double Insulated Product ETL certified complying with UL1431 and UL1097 ETL certified to CSA C22.2 No. 68. No extension cords. Part No 1143482 13 Perfecto2™Series