Invacare IRC5PO2V Owners Manual - Page 21

Front Panel Indicators - Units without, SensO - concentrator

Page 21 highlights

SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum. LABEL SYMBOL O2 O2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O2 over 85% (± 2%) O2 Between 73% (± 3%) to 85% (±2%) SYSTEM FAILURE O2 Below 73% (±3%) GREEN Indicator Light YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualified technician. Have a back-up supply of oxygen ready. RED Indicator Light Continuous Audible Alarm Sieve-GARD Compressor Shutdown. Call a qualified technician. Switch to a back-up supply of oxygen immediately. Front Panel Indicators - Units without SensO2 NOTE: For this procedure, refer to FIGURE 5.7. Auto Shut Down Use Backup RED Call Supplier Normal I/O GREEN FIGURE 5.7 Front Panel Indicators - Units without SensO2 Part No 1143482 21 Perfecto2™Series

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

SECTION 5—OPERATING INSTRUCTIONS
Part No 1143482
21
Perfecto
2
Series
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm
up
time (approximately 30 min.) while waiting for the O
2
purity to
reach maximum.
Front Panel Indicators - Units without
SensO
2
NOTE: For this procedure, refer to FIGURE 5.7.
FIGURE 5.7
Front Panel Indicators - Units without SensO2
LABEL
SYMBOL
O
2
PURITY
INDICATOR LIGHTS (LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician. Have a
back-up supply of oxygen ready.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down. Call a qualified technician.
Switch to a back-up supply of oxy-
gen immediately.
O
2
I/O
Auto Shut Down
Use Backup
Normal
RED
GREEN
Call Supplier