Invacare POC1-100B Owners Manual - Page 59

System Fault

Page 59 highlights

Troubleshooting LCD DISPLAY TEXT: DESCRIPTION: SOLUTIONS: Shut Down System Fault INDICATORS: Two audible beeps every 16 seconds Abnormal system condition has been detected. 1. Turn the concentrator Off, then On again May be caused by malfunction of compressor, to retry. fan, pressure or software. 2. Change to another source of oxygen if alarm continues Concentrator function has shut down. YELLOW alarm indicator illuminated NOTES: To clear the alarm, turn the concentrator off. Restart concentrator. Call your provider and report the problem if alarm continues. LCD DISPLAY TEXT: DESCRIPTION: Not applicable. All power lost, display is blank. All power sources have been lost/removed. INDICATORS: Concentrator function has shut down. Audible beeps every second for approximately 30 to 60 seconds No other control panel indicators illuminated NOTES: This alarm occurs even if the concentrator is off when all power is removed. To clear the alarm, complete solutions 1, 2, or 3 before restarting. SOLUTIONS: 1. Ensure the battery is completely inserted. 2. Check external power connections to the concentrator and to the power outlet. 3. Try another power outlet. 4. Change to another source of oxygen if alarm continues. 1194969-A-00 59

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72

Troubleshooting
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Shut Down
System Fault
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
Abnormal system condition has been detected.
May be caused by malfunction of compressor,
fan, pressure or software.
Concentrator function has shut down.
1.
Turn the concentrator Off, then On again
to retry.
2.
Change to another source of oxygen if
alarm continues
NOTES:
To clear the alarm, turn the concentrator off. Restart concentrator.
Call your provider and report the problem if alarm continues.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Not applicable. All power lost, display is blank.
INDICATORS:
Audible beeps every second for approximately
30 to 60 seconds
No other control panel indicators illuminated
All power sources have been lost/removed.
Concentrator function has shut down.
1.
Ensure the battery is completely inserted.
2.
Check external power connections to the
concentrator and to the power outlet.
3.
Try another power outlet.
4.
Change to another source of oxygen if
alarm continues.
NOTES:
This alarm occurs even if the concentrator is off when all power is removed.
To clear the alarm, complete solutions 1, 2, or 3 before restarting.
1194969-A-00
59