Invacare POC1-100B Owners Manual - Page 59
System Fault
View all Invacare POC1-100B manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 59 highlights
Troubleshooting LCD DISPLAY TEXT: DESCRIPTION: SOLUTIONS: Shut Down System Fault INDICATORS: Two audible beeps every 16 seconds Abnormal system condition has been detected. 1. Turn the concentrator Off, then On again May be caused by malfunction of compressor, to retry. fan, pressure or software. 2. Change to another source of oxygen if alarm continues Concentrator function has shut down. YELLOW alarm indicator illuminated NOTES: To clear the alarm, turn the concentrator off. Restart concentrator. Call your provider and report the problem if alarm continues. LCD DISPLAY TEXT: DESCRIPTION: Not applicable. All power lost, display is blank. All power sources have been lost/removed. INDICATORS: Concentrator function has shut down. Audible beeps every second for approximately 30 to 60 seconds No other control panel indicators illuminated NOTES: This alarm occurs even if the concentrator is off when all power is removed. To clear the alarm, complete solutions 1, 2, or 3 before restarting. SOLUTIONS: 1. Ensure the battery is completely inserted. 2. Check external power connections to the concentrator and to the power outlet. 3. Try another power outlet. 4. Change to another source of oxygen if alarm continues. 1194969-A-00 59