Invacare POC1-100B Owners Manual - Page 66

Invacare POC1-100B Manual

Page 66 highlights

Invacare® Platinum™ Mobile Surge IEC 61000-4-5 ± 1 kV line(s) to line(s) ± 2 kV line(s) to ground Voltage dips IEC 61000-4-11 0% UT ; 1 cycle and 70% UT ; 25/30 cycles Single phase: at 0° Voltage Interruptions 0% UT ; 250/300 cycle IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 30 A/m ± 1 kV line(s) to line(s) ± 2 kV line(s) to ground 0% UT ; 1 cycle and 70% UT ; 25/30 cycles Single phase: at 0° 0% UT ; 250/300 cycle 30 A/m Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the Device requires continued operation during power mains interruptions, it is recommended that the Device be powered from an un-interruptible power supply or a battery. UT is the a. c. mains voltage prior to application of the test level. Mains power quality should be that of a typical commercial or hospital environment. If the user of the Device requires continued operation during power mains interruptions, it is recommended that the Device be powered from an un-interruptible power supply or a battery. UT is the a. c. mains voltage prior to application of the test level. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. 66 1194969-A-00

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72

Invacare® Platinum™ Mobile
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips
IEC 61000-4-11
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30 cycles
Single phase: at 0°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30 cycles
Single phase: at 0°
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Device requires
continued operation during power mains interruptions, it
is recommended that the Device be powered from an
un-interruptible power supply or a battery.
U
T
is the a. c. mains voltage prior to application of the
test level.
Voltage Interruptions
IEC 61000-4-11
0%
U
T
; 250/300 cycle
0%
U
T
; 250/300 cycle
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Device requires
continued operation during power mains interruptions, it
is recommended that the Device be powered from an
un-interruptible power supply or a battery.
U
T
is the a. c. mains voltage prior to application of the
test level.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
66
1194969-A-00