Invacare RPS350-1 Owners Manual - Page 20
Troubleshooting - actuator
View all Invacare RPS350-1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
SECTION 6-TROUBLESHOOTING SECTION 6-TROUBLESHOOTING SYMPTOMS Noisy or dry sound from pivots. Electric actuator fails to lift when button is pressed. FAULTS Needs lubrication. Hand-control or actuator connector loose. Battery low. RED emergency stop button pressed IN. SOLUTION Refer to Lubricating the Lift on page 22. Check the hand control and actuator connections. Re-connect as necessary. Charge batteries. Refer to Charging the Battery on page 13. Rotate RED emergency stop button CLOCKWISE until it pops out. Battery not connected properly Reconnect the battery to the control box. to control box. Refer to Charging the Battery on page 13. The connecting terminals are damaged. Replace the battery pack. Refer to Charging the Battery on page 13. Electric actuator in need of ser- Refer to Adjusting the Base on page 22. vice or load is too high. Contact Dealer if service is required. Unusual noise from actuator. Actuator is worn or damaged or spindle is bent. Replace the actuator or contact Dealer. Refer to Replacing the Electric Actuator on page 23. Lift arms will not lower from the uppermost position. Lift arms require a minimum weight load to lower from the uppermost position. Pull down slightly on the lift arms. NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare. Stand Up Patient Lift 20 Part No. 1078984