Kodak I1860 User Guide - Page 43

UDDS Calibration, Calibrate, Do not use the calibration target, End Job, paper in the input elevator.

Page 43 highlights

UDDS Calibration UDDS calibration is used to calibrate the ultrasonic document detection system for multi-feeds. UDDS calibration is rarely needed. This calibration should only be performed if there is a feeding or multi-feed detection issue. 1. Touch Calibrate. The Calibration screen will be displayed. 2. Touch UDDS Calibration to display the UDDS Calibration screen. 3. You will be prompted to place a 20 lb. or 75 g/m2 bond sheet of paper in the input elevator. Do not use the calibration target. 4. Touch OK. The calibration results will be displayed. 5. When finished, touch End Job to return to the Idle screen. 3-26 A-61555 April 2009

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108

3-26
A-61555
April 2009
UDDS Calibration
UDDS calibration is used to calibrate the ultrasonic document detection
system for multi-feeds. UDDS calibration is rarely needed. This
calibration should only be performed if there is a feeding or multi-feed
detection issue.
1. Touch
Calibrate
. The Calibration screen will be displayed.
2. Touch
UDDS Calibration
to display the UDDS Calibration screen.
3.
You will be prompted to place a 20 lb. or 75 g/m
2
bond sheet of
paper in the input elevator.
Do not use the calibration target
.
4. Touch
OK
. The calibration results will be displayed.
5.
When finished, touch
End Job
to return to the Idle screen.