Configuring the CWS655
![]() |
View all Campbell Scientific CWS220 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 31 highlights
Wireless Sensor Network 3.3.2 Configuring the CWS655 To configure the CWS655 Water Content Reflectometer, first connect the A205 to the type B USB connector on the 17648 cable. Remove the battery door on the CWS655 and locate the four-pin connector near the top of the sensor. Connect the A205 to the sensor at the four-pin connector. The A205 label should be facing toward the CWS655 antenna. Select the Program settings for CWS655_1:CWS655 item in the Configure Devices list, then select Click Here. When the Setup Dialog window for configuring the CWS655 opens, verify that the Serial Port matches the port number where the A205 is connected and click Connect. Press the Setup button on the back of the CWS655. Settings will be transferred and applied to the CWS655. In the Settings have Been Applied window, options are available for saving the settings to a XML file, printing the settings, and comparing the settings to another XML file. When finished with these actions, click OK. NOTE The Setup button must be pressed to enable communications between the wireless sensor and computer. 23