Campbell Scientific LI200X LI200X-L LI-COR Silicon Pyranometer - Page 6
Installation
View all Campbell Scientific LI200X manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 6 highlights
LI200X Pyranometer NOTE The black outer jacket of the cable is Santoprene® rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings. LP R I COR YR PY A N 2 O 0 M 0 E 2 T 3 E FIGURE 1-1. LI200X Pyranometer 2. Installation The LI200X should be mounted such that it is never shaded by the tripod/tower or other instrumentation. The sensor should be mounted with the cable pointed towards the nearest magnetic pole, e.g. in the Northern Hemisphere point the cable towards the North Pole. Mounting height is not critical for the accuracy of the measurement. However, pyranometers mounted at heights of 3 m or less are easier to level and clean. To ensure accurate measurements, the LI200X should be mounted using LI2003S base/leveling fixture. This base incorporates a bubble level and three adjustment screws. The LI200X and base/leveling fixture are attached to a tripod or tower using one of three mounting configurations (see Figure 2-1 through 2-3). Tools required for installation on a tripod or tower: Small and medium Phillips screwdrivers 5/32" Allen wrench for NU-RAIL (Figure 2-3) 1/2" open end wrench for 015ARM or CM225 (Figures 2-1, 2-2) Tape measure UV-resistant wire ties Side-cut pliers Compass Step ladder 2