Campbell Scientific PVS4151C PVS4100/4120/4150 Portable Automatic Samplers - Page 58
ARROW, Interval Time, ENTER, SAMPLE INITIATION, RESTART, Summary of Sequence, RESTART, RESTART
![]() |
View all Campbell Scientific PVS4151C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 58 highlights
Section 5. Programming 3. Press an ARROW key. Continue until Interval Time is shown on the display. 4. Press the ENTER key. When setting the time, a flashing digit prompts for input from a numeric key. To set the time, press a numeric key to enter a value and advance to each digit in succession. The format is HHH:MM. The minimum time can be set to 1 minute, however, practical considerations, such as equipment duty cycle, maintenance and service life suggest times of 3 minutes or longer. When the five digits are entered, press ENTER. Any wrong entries will require re-entry. There are two methods of correcting a mistake. The digit flashing "wraps around" and begins again, at which time the correct entry may be pressed. Alternately, the arrow keys can be used to reposition the flashing prompt over the error, which can then be replaced with the correct value. The display will echo the last entry with . The controller has now been given a new value for SAMPLE INITIATION. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait the designated time before taking a sample. Summary of Sequence: SET, SAMPLE INITIATION, ARROW(S), ENTER, ###:##, ENTER, RESTART, RESTART. 5-18