Campbell Scientific SR50A SR50A Distance Sensor - Page 14
RS-232 and RS-485 Settings
View all Campbell Scientific SR50A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
SR50A Sonic Ranging Sensor FIGURE 1. SR50A to MD485 Wiring 3.6 RS-232 and RS-485 Settings Once the jumpers are set for RS-232 operation a terminal program such as Hyperterminal can be used to change factory default or existing settings. The following settings apply to Hyperterminal or any other program that is used for communications. BAUD Rate Data Bits Parity Stop Bits Flow Control Current SR50A SettingNote 8 None 1 None NOTE The factory default BAUD rate is 9600 BAUD. Once the BAUD rate is changed the new BAUD rate must be used for further communications to the SR50A. It is important to keep track of the BAUD rate setting on the SR50A. If the BAUD rate setting is unknown, we recommend trying the default value of 9600 BAUD. If that does not work, start at BAUD rate 1200 and go through all the BAUD rate settings until the correct one is found. When the hyperterminal session is started the command "SETUP" needs to be entered. Hitting the Enter will transmit the CR LF characters which are required after the text "SETUP". The text "SETUP" is not case sensitive so any combination of upper and lower case letters will work. 8