Fluke 116 Fluke 116 Users Manual - Page 14

Measuring AC and DC Current - meters

Page 14 highlights

116 Users Manual Measuring AC and DC Current WWarning To avoid personal injury or damage to the Meter: • Never attempt to make an in-circuit current measurement when the opencircuit potential to earth is > 600 V. • Use the proper switch position and range for your measurement. Flame sensor probe Control Module eeo08f.eps To measure flame rectification circuits: 1. Turn the function switch to $. 2. Connect the Meter between the flame sensor probe and the control module. 3. Turn heating unit on and record μA measurement. 12

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

116
Users Manual
12
Measuring AC and DC Current
W
Warning
To avoid personal injury or damage to the
Meter:
Never attempt to make an in-circuit
current measurement when the open-
circuit potential to earth is
>
600 V.
Use the proper switch position and range
for your measurement.
Flame
sensor
probe
Control
Module
eeo08f.eps
To measure flame rectification circuits:
1.
Turn the function switch to
$
.
2.
Connect the Meter between the flame sensor probe and
the control module.
3.
Turn heating unit on and record
μ
A measurement.