Fluke 1577 Fluke 1577 and 1587 Insulation Multimeter Datasheet - Page 2

Specifications

Page 2 highlights

1587/1577 Specifications AC voltage measurement 1587 accuracy Range 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V 11 kHz bandwidth Resolution 0.1 mV 0.001 V 0.01 V 0.1 V 1 V 50 Hz to 60 Hz + (1 % + 3) + (1 % + 3) + (1 % + 3) + (1 % + 3) + (2 % + 3) 60 Hz to 5000 Hz + (2 % + 3) + (2 % + 3) + (2 % + 3) + (2 % + 3)1 + (2 % + 3)1 1587 lowpass filter Voltage Range 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V Resolution 0.1 mV 0.001 V 0.01 V 0.1 V 1 V 50 Hz to 60 Hz + (1 % + 3) + (1 % + 3) + (1 % + 3) + (1 % + 3) + (2 % + 3) 60 Hz to 400 Hz + (2 % + 3) + (2 % + 3) + (2 % + 3) + (2 % + 3) + (2 % + 3) 1577 accuracy Range 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V Resolution 0.1 mV 0.001 V 0.01 V 0.1 V 1 V 50 Hz to 60 Hz + (2 % + 3) + (2 % + 3) + (2 % + 3) + (2 % + 3) + (2 % + 3) AC conversion: Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at full scale up to 500 V, decreasing linearly to crest factor < = 1.5 at 1000 V. For non-sinusoidal waveforms add ± (2 % reading + 2 % FS) typical, for a crest factor up to 3. Input impedance: 10 MΩ (nominal), < 100 pF, ac-coupled Common mode rejection ratio (1 kΩ unbalanced): > 60 dB at dc, 50 or 60 Hz Overload protection: 1000 V rms or dc, 107 V Hz Max DC voltage measurement Range 6.000 V dc 60.00 V dc 600.0 V dc 1000 V dc 1Accuracies apply to ± 100 % of range Resolution 0.001 V 0.01 V 0.1 V 1 V Accuracy 15871 0.09 % + 2 0.09 % + 2 0.09 % + 2 0.09 % + 2 Input impedance: 10 MΩ (nominal), < 100 pF Normal mode rejection ratio: > 60 dB @ 50 Hz or 60 Hz Common rode rejection ratio: > 120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) Overload protection: 1000 V rms or dc Accuracy 15771 0.2 % + 2 0.2 % + 2 0.2 % + 2 0.2 % + 2 DC millivolts measurement Range 600.0 mV dc Resolution 0.1 mV Accuracy 1587 0.1 % + 1 Accuracy 1577 0.2 % + 1 2 Fluke Corporation 1587/1577 Insulation Multimeters

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6

2
Fluke Corporation
1587/1577 Insulation Multimeters
AC voltage measurement
1587 accuracy
Range
Resolution
50 Hz to 60 Hz
60 Hz to 5000 Hz
600.0 mV
0.1 mV
+ (1 % + 3)
+ (2 % + 3)
6.000 V
0.001 V
+ (1 % + 3)
+ (2 % + 3)
60.00 V
0.01 V
+ (1 % + 3)
+ (2 % + 3)
600.0 V
0.1 V
+ (1 % + 3)
+ (2 % + 3)
1
1000 V
1 V
+ (2 % + 3)
+ (2 % + 3)
1
1
1 kHz bandwidth
1587 lowpass filter Voltage
Range
Resolution
50 Hz to 60 Hz
60 Hz to 400 Hz
600.0 mV
0.1 mV
+ (1 % + 3)
+ (2 % + 3)
6.000 V
0.001 V
+ (1 % + 3)
+ (2 % + 3)
60.00 V
0.01 V
+ (1 % + 3)
+ (2 % + 3)
600.0 V
0.1 V
+ (1 % + 3)
+ (2 % + 3)
1000 V
1 V
+ (2 % + 3)
+ (2 % + 3)
1577 accuracy
Range
Resolution
50 Hz to 60 Hz
600.0 mV
0.1 mV
+ (2 % + 3)
6.000 V
0.001 V
+ (2 % + 3)
60.00 V
0.01 V
+ (2 % + 3)
600.0 V
0.1 V
+ (2 % + 3)
1000 V
1 V
+ (2 % + 3)
AC conversion:
Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are
true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at
full scale up to 500 V, decreasing linearly to crest factor < = 1.5 at 1000 V. For non-sinusoidal waveforms
add ± (2 % reading + 2 % FS) typical, for a crest factor up to 3.
Input impedance:
10 M
Ω
(nominal), < 100 pF, ac-coupled
Common mode rejection ratio (1 k
Ω
unbalanced):
> 60 dB at dc, 50 or 60 Hz
Overload protection:
1000 V rms or dc, 10
7
V Hz Max
DC voltage measurement
Range
Resolution
Accuracy 1587
1
Accuracy 1577
1
6.000 V dc
0.001 V
0.09 % + 2
0.2 % + 2
60.00 V dc
0.01 V
0.09 % + 2
0.2 % + 2
600.0 V dc
0.1 V
0.09 % + 2
0.2 % + 2
1000 V dc
1 V
0.09 % + 2
0.2 % + 2
1
Accuracies apply to ± 100 % of range
Input impedance:
10 M
Ω
(nominal), < 100 pF
Normal mode rejection ratio:
> 60 dB @ 50 Hz or 60 Hz
Common rode rejection ratio:
> 120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)
Overload protection:
1000 V rms or dc
DC millivolts measurement
Range
Resolution
Accuracy 1587
Accuracy 1577
600.0 mV dc
0.1 mV
0.1 % + 1
0.2 % + 1
1587/1577 Specifications