Fluke 1623-2 Manual - Page 38

Specifications, RA 3-Pole and 4-Pole ground resistance measurement

Page 38 highlights

1623-2 Users Manual Specifications Temperature ranges Operating temperature range: Storage temperature range: Temperature coefficient: Operating humidity: Operating altitude: Climatic class: Protection type Case: Battery door: Electromagnetic compatibility: 0 °C to +35 °C (+32 °F to +95 °F) -20 °C to +60 °C (-4 °F to +140 °F) ±0.1 % of rdg / °C (below 18 °C and above 28 °C) 120 dB (162/3, 50, 60, 400 Hz) 6 seconds, typical 250 Vrms (pertains to misuse) 6 x 1.5 v, AA, LR6 Alkaline >3000 measurements, typical 240 mm x 180 mm x 110 mm (9.5 in x 7.1 in x 4.4 in) 1.49 kg (3.28 lb) Internal memory storage up to 1500 records accessible via USB port RA 3-Pole and 4-Pole ground resistance measurement Resolution Measurement range Accuracy Operating error 0.001 ... 10 Ω 0.020 Ω to 19,99 kΩ ±(2 % rdg + 3 d) ±(5 % rdg + 3 d) Note For 2-pole measurements, connect terminals H and S with the supplied connector cable. 30 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

1623-2
Users Manual
30
Specifications
Temperature ranges
Operating temperature range:
0 °C to +35 °C (+32 °F to +95 °F)
Storage temperature range:
-20 °C to +60 °C (-4 °F to +140 °F)
Temperature coefficient:
±
0.1 % of rdg / °C (below 18 °C and above 28 °C)
Operating humidity:
<95 % RH noncondensing
Operating altitude:
2000 m
Climatic class:
C1 (IEC 654-1), -5 °C to +45 °C, 5 % to 95 % RH
Protection type
Case:
IP56
Battery door:
IP40
Electromagnetic compatibility:
Complies with IEC61326-1: Portable
Safety:
Complies with IEC 61010-1: CAT None, Pollution Degree 2
External voltage:
U
ext
, max = 24 V (dc, ac < 400 Hz), measurement inhibited
for higher values
Noise rejection:
>120 dB (162/3, 50, 60, 400 Hz)
Measurement time:
6 seconds, typical
Maximum overload:
250 V
rms
(pertains to misuse)
Batteries:
6 x 1.5 v, AA, LR6 Alkaline
Battery life span:
>3000 measurements, typical
Dimensions:
240 mm x 180 mm x 110 mm (9.5 in x 7.1 in x 4.4 in)
Weight with batteries:
1.49 kg (3.28 lb)
Memory:
Internal memory storage up to 1500 records
accessible via USB port
RA 3-Pole and 4-Pole ground resistance measurement
Resolution
Measurement range
Accuracy
Operating error
0.001 … 10
Ω
0.020
Ω
to 19,99 k
Ω
±
(2 % rdg + 3 d)
±
(5 % rdg + 3 d)
Note
For 2-pole measurements, connect terminals H and S with the supplied
connector cable.
Shop for Fluke products online at:
1.877.766.5412
www.
MyFlukeStore
.com