Fluke 1623 FE 1623 Users Manual - Page 17
RA 4-pole Measurements, Refer - 2 earth ground tester
View all Fluke 1623 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 17 highlights
Earth/Ground Tester Setup RA 4-pole Measurements Refer to Figure 5. A Select function RA 4-pole. Display as shown below. B Connect test leads. Connect terminals E/C1 and ES/P1 to the earth system to be measured with the 2 supplied test leads (1.5 m). Place 2 ground stakes in earth/dirt. Minimum distance between earth electrode (E/C1), probe (S/P2), and auxiliary earth (H/C2) should be at least 20 m! The ES test lead eliminates the influence of the test leads. Connect the stakes with the 25 m and 50 m cable reels to H/C2 and S/P2 as shown below. C Press START. The "active" symbol indicates that a measurement is in progress. For a continuous measurement keep START pressed. D The symbol "9" indicates a completed measurement. The result is kept on the display until a new measurement is started or the rotary switch is turned. 1 2 E ES S H FLUKE 1623 Earth/Ground Tester max 50 V P2 S P1 ES/P1 ES E/C1 S/P2 H/C2 plus FLUKE E >20 m >20 m 3 4 active E ES S H E ES S H Figure 5. R 4-pole Measurements A edv004.eps 9