Fluke 1630-2 FC FE 1630 Users Manual - Page 15
Memory and Alarm Functions, Using the High and Low Alarms
View all Fluke 1630-2 FC manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
Earth Ground Clamp Memory and Alarm Functions Memory and Alarm Functions Using the High and Low Alarms 1. Press F to select the HI or LO alarm, although this describes the ohms function high and low alarms can be set in any of the other functions. The current value of High or Low alarm will appear on the display. Figure 5 shows alarm function displays. 2. Press the E or E button to increment/decrement the value by 1 Ω. The value can be incremented or decremented from 0 Ω to 1510 Ω and then OL. Hold down the up or down buttons to quickly increment/decrement to the desired value. 3. When the value is set, press F until you return to the main display. 4. When the rotary switch is set to the U position, the Clamp compares the current value with the high and low values. If the current measurement is larger than the HI value, the Clamp beeps and shows HI-- in the upper display. If the measurement is less than the LO value, the Clamp will beep and LO-- appears in the upper display. Figure 5. Alarm Functions eht06.eps 9