Fluke 175 FE 175,177,179 Users Manual - Page 18
Function, Overload Protection, Overload, Full Scale Voltage
View all Fluke 175 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 18 highlights
Models 175, 177 & 179 Users Manual Function Volts AC Volts DC mV/T Overload Protection 1 1000 V RMS 1000 V RMS 1000 V RMS2 Input Impedance (Nominal) > 10 MΩ < 100 pF > 10 MΩ 10 MΩ < 100 pF Open Circuit Test Voltage Ohms 1000 V RMS2 Continuity/Diode test 1000 V RMS2 1. 10 7 V-Hz maximum. < 8.0 V DC < 8.0 V DC 2. For circuits < 0.3 A short circuit. 660 V for high energy circuits. Common Mode Rejection Ratio (1 kΩ Unbalanced) > 60 dB @ DC, 50 Hz or 60 Hz >120 dB @ DC, 50 Hz or 60 Hz >120 dB @ DC, 50 Hz or 60 Hz Full Scale Voltage To: 600 kΩ 50 MΩ < 660 mV DC < 4.6 V DC 2.4 V DC Normal Mode Rejection > 60 dB @ 50 Hz or 60 Hz > 60 dB @ 50 Hz or 60 Hz Short Circuit Current < 1.1 mA < 1.1 mA Function mA A Overload Protection Fused, 44/100 A, 1000 V FAST Fuse Fused, 11 A, 1000 V FAST Fuse Overload 600 mA overload for 2 minutes maximum 20 A overload for 30 seconds maximum 14