Fluke 179 FE 175,177,179 Users Manual - Page 15
Specifications, ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage Category III
View all Fluke 179 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
Specifications Specifications Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 95 %. Accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] ) Maximum voltage between any terminal and earth ground: 1000 V DC or AC RMS Surge Protection: W Fuse for mA inputs: W Fuse for A input: 8 kV peak per IEC 61010 440 mA, 1000 V FAST Fuse 11 A, 1000 V FAST Fuse Display: Digital: 6000 counts, updates 4/sec Bar Graph: 33 segments; Updates 40/sec Frequency: 10,000 counts Capacitance: 1,000 counts Altitude: Operating: 2000 m; Storage: 12,000 m Temperature: Operating: −10 °C to +50 °C; Storage: −30 °C to +60 °C Temperature coefficient: 0.1 X (specified accuracy / °C (< 18 °C or > 28 °C) Electromagnetic Compatibility (EN 61326-1:1997): In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: specified accuracy ± 5 °C (9 °F) Relative Humidity: Noncondensing < 0 °C 0 % to 95 % @ 10 °C to 30 °C 0 % to 75 % @ 30 °C to 40 °C; 0 % to 40 % @ 40 °C to 50 °C Battery Life: Alkaline: 300 hrs typical Size (H x W x L): 4.3 cm x 9 cm x 18.5 cm Weight: 420 g Safety Compliances: Certifications: ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage Category III, 600 V Overvoltage P ; Category IV CSA, TÜV (EN61010), UL, , (N10140),VDE 11