Fluke 2638A/05/C 120 Product Manual - Page 102

Troubleshooting, Table 7-2. Troubleshooting Chart, Problem, Causes and Solutions, Cause 1, Solution 1

Page 102 highlights

7 Error Messages and Troubleshooting Troubleshooting Troubleshooting In the event that the Product functions abnormally, Table 7-2 can help find and solve the problem. Several possible problem conditions are described with likely causes and solutions. If a problem occurs, please read this section carefully and attempt to understand and solve the problem. If the Product seems faulty or the problem cannot otherwise be solved, contact an Authorized Service Center for assistance, see Chapter 1. Be sure to have the instrument model number, serial number, and regional voltage available to tell the Service Technician. Table 7-2. Troubleshooting Chart Problem The Product does not power on. LED lights are illuminated but the display is blank. Input Module is not recognized. Causes and Solutions Cause 1: Fuses. Solution 1: If a fuse blows, it could be because of a power surge or a component failure. Replace the fuse once. DO NOT replace the fuse with one of a higher current rating. Always replace the fuse with one of the same rating, voltage, and type. If the fuse blows a second time, it is likely caused by failure of a component part. Cause 1: Power Cord. Solution 1: Check that the power cord is plugged in and connected to the Product. Cause 2: AC mains power. Solution 2: Make sure the circuit that supplies power to the Product is on. Cause 3: Power entry is not properly set. Solution 3: Make sure the power entry setting is matching with the AC mains power supply rate. Cause 1: The unit is standby mode (the power LED should be amber). Solution 1: Press the Standby key on the front panel to exit the standby mode. Cause 2: Screen saver activated. Solution 2: Press any front panel key to exit the screen saver. Cause 3: Screen or unit is malfunctioning. Solution 3: Contact Fluke Calibration. See Chapter 1. Cause 1: Input Module not fully seated. Solution 1: Reinstall the Input Module and make sure it is fully seated in the slot. Cause 2: Relay card not installed. Solution 2: Make sure the relay card is installed. Cause 3: Input Module has failed. Solution 3: Contact an Authorized Customer Service Center. 7-17 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103

Error Messages and Troubleshooting
Troubleshooting
7
7-17
Troubleshooting
In the event that the Product functions abnormally, Table 7-2 can help find and solve the
problem. Several possible problem conditions are described with likely causes and
solutions. If a problem occurs, please read this section carefully and attempt to
understand and solve the problem. If the Product seems faulty or the problem cannot
otherwise be solved, contact an Authorized Service Center for assistance, see Chapter 1.
Be sure to have the instrument model number, serial number, and regional voltage
available to tell the Service Technician.
Table 7-2. Troubleshooting Chart
Problem
Causes and Solutions
The Product does not
power on.
Cause 1:
Fuses.
Solution 1:
If a fuse blows, it could be because of a power surge or a
component failure. Replace the fuse once. DO NOT replace the fuse with one
of a higher current rating. Always replace the fuse with one of the same rating,
voltage, and type. If the fuse blows a second time, it is likely caused by failure
of a component part.
Cause 1:
Power Cord.
Solution 1:
Check that the power cord is plugged in and connected to the
Product.
Cause 2:
AC mains power.
Solution 2:
Make sure the circuit that supplies power to the Product is on.
Cause 3:
Power entry is not properly set.
Solution 3:
Make sure the power entry setting is matching with the AC mains
power supply rate.
LED lights are
illuminated but the
display is blank.
Cause 1:
The unit is standby mode (the power LED should be amber).
Solution 1:
Press the Standby key on the front panel to exit the standby
mode.
Cause 2:
Screen saver activated.
Solution 2:
Press any front panel key to exit the screen saver.
Cause 3:
Screen or unit is malfunctioning.
Solution 3:
Contact Fluke Calibration. See Chapter 1.
Input Module is not
recognized.
Cause 1:
Input Module not fully seated.
Solution 1:
Reinstall the Input Module and make sure it is fully seated in the
slot.
Cause 2:
Relay card not installed.
Solution 2:
Make sure the relay card is installed.
Cause 3:
Input Module has failed.
Solution 3:
Contact an Authorized Customer Service Center.
Shop for Fluke products online at:
1.877.766.5412
www.
MyFlukeStore
.com