Fluke 2638A/20 Product Manual - Page 63

Digital I/O (DIO) Channel Configuration (Ch401), digital I/O DIO port.

Page 63 highlights

3 Input and Channel Configuration Channel Configuration Digital I/O (DIO) Channel Configuration (Ch401) The Product can sense and output a digital, 8-bit transistor-transistor logic (TTL) value that can be displayed as the 8-bit TTL value and be recorded as the decimal equivalent. This is accomplished with the rear-panel digital I/O connectors (DIO) that are shown in Figure 3-7. When a DIO input is sensed, the Product displays the status of the digital I/O as 8-bit TTL value and records the decimal equivalent. The 8-bit TTL value can also be seen when a channel is monitored. The default value is a high state that shows as "11111111" until the Product senses a low state or is used as an output. A few examples of some 8-bit TTL values and their decimal equivalents are shown below: • 11111111 is represented by decimal 255 • 00001111 is represented by decimal 15 • 00010001 is represented by decimal 17 • 10000101 is represented by decimal 133 Notes For input and output specifications such as input and output voltages, see "Digital I/O" in Chapter 1. The DIO channel will be read-only when it is set to ON (active). With a remote command, the Product can output an 8-bit TTL value. See the 2638A Remote Programmers Guide for more information. Terminal TOT 1 2 3 4 5 6 7 8 GND Function Totalizer Input Input/Output Line 1 Input/Output Line 2 Input/Output Line 3 Input/Output Line 4 Input/Output Line 5 Input/Output Line 6 Input/Output Line 7 Input/Output Line 8 Ground Terminal Figure 3-7. DIO Connector hcn040.eps Set up the DIO channel as follows: 1. Connect the equipment to the DIO input terminal then insert it into the rear-panel digital I/O (DIO) port. 2. Push . 3-23 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103

Input and Channel Configuration
Channel Configuration
3
3-23
Digital I/O (DIO) Channel Configuration (Ch401)
The Product can sense and output a digital, 8-bit transistor-transistor logic (TTL) value
that can be displayed as the 8-bit TTL value and be recorded as the decimal equivalent.
This is accomplished with the rear-panel digital I/O connectors (DIO) that are shown in
Figure 3-7.
When a DIO input is sensed, the Product displays the status of the digital I/O as 8-bit
TTL value and records the decimal equivalent. The 8-bit TTL value can also be seen
when a channel is monitored. The default value is a high state that shows as “11111111”
until the Product senses a low state or is used as an output. A few examples of
some 8-bit
TTL values and their decimal equivalents are shown below:
11111111 is represented by decimal 255
00001111 is represented by decimal 15
00010001 is represented by decimal 17
10000101 is represented by decimal 133
Notes
For input and output specifications such as input and output voltages, see
“Digital I/O” in Chapter 1.
The DIO channel will be read-only when it is set to ON (active).
With a remote command, the Product can output an 8-bit TTL value. See the
2638A Remote Programmers Guide
for more information.
Terminal
TOT
1
2
3
4
5
6
7
8
GND
Function
Totalizer Input
Input/Output Line 1
Input/Output Line 2
Input/Output Line 3
Input/Output Line 4
Input/Output Line 5
Input/Output Line 6
Input/Output Line 7
Input/Output Line 8
Ground Terminal
hcn040.eps
Figure 3-7. DIO Connector
Set up the DIO channel as follows:
1.
Connect the equipment to the DIO input terminal then insert it into the rear-panel
digital I/O (DIO) port.
2.