Illustration of a Scan Sweep, Table 4-2. Settling Time Delay Values, Type and Range
View all Fluke 2638A/20 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 77 highlights
2638A Users Manual • The total channel sample time is the minimum sample time plus the userprogrammed channel delay. The longer the channel delay, the longer it takes to sample the channel. • The total scan sweep time is sum of the channel sample times of all the channels scanned. A scan sweep will start when the selected trigger source is asserted if a previous scan sweep is not still in progress (see "Trigger Type" on page 4-6). • The interval at which scan sweeps occur can be fixed by programming the trigger type to Timer and setting an interval time. SCAN START Channel List Channel 102 Channel 104 Key: - Sample Time* - Channel Delay *Consists of NPLC time, settling time, and ADC samples. Scan Cycle 1 Remaining Interval Scan Cycle 2 RReemmaaiinniinngg IInntteerrvvaall Channel 115 Channel Delay Set Channel 117 Channel Delay Set Channel 121 Sweep Time Scan Interval Sweep Time Scan Interval SCAN FINISH Figure 4-2. Illustration of a Scan Sweep Table 4-2. Settling Time Delay Values Type and Range 100 mV dc 1 V dc 10 V dc 100 V dc 300 V dc ACV all ranges 20 Hz filter ACV all ranges 200 Hz filter A dc (all ranges) Time Delay 0 ms 0 ms 0 ms 100 ms 100 ms 400 ms 100 ms 0 ms hcn050.eps 4-4 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664