AC Current Specifications Function Range Resolution AC μA [3] 500 μA 0.01 μA AC mA [3] 5000 μA 50 mA 0.1 μA 0.001 mA AC A [2] 400 mA 5 A 10 A [1] 0.01 mA 0.0001 A 0.001 A [1] 10 A range (10 % to 100 % of range). [2] 20 A for 30 seconds on, 10 minutes off. >10 A not specified. [3] 400 mA continuous; 550 mA for 2 minutes on, 1 minute off. [4] Verified by design and type tests. See Detailed Specifications introduction for additional information. 20 to 45 Hz 1 % + 20 1 % + 5 1 % + 20 1 % + 5 1.5 % + 20 1.5 % + 5 True-rms Digital Multimeters Detailed Specifications Accuracy 45 to 1 kHz 1 to 20 kHz 0.6 % + 20 0.6 % + 20 0.6 % + 5 0.6 % + 10 0.6 % + 20 0.6 % + 20 0.6 % + 5 0.8 % + 20 0.8 % + 5 1.5 % + 10 3 % + 40 [4] 3 % + 10 [4] 20 to 100 kHz [4] 5 % + 40 5 % + 40 5 % + 40 5 % + 40 Not Spec'd Not Spec'd 69

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88

True-rms Digital Multimeters
Detailed Specifications
69
AC Current Specifications
Accuracy
Function
Range
Resolution
20 to 45 Hz
45 to 1 kHz
1 to 20 kHz
20 to 100 kHz
[4]
500
μ
A
0.01
μ
A
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC
μ
A
[3]
5000
μ
A
0.1
μ
A
1 % + 5
0.6 % + 5
0.6 % + 10
5 % + 40
50 mA
0.001 mA
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC mA
[3]
400 mA
0.01 mA
1 % + 5
0.6 % + 5
1.5 % + 10
5 % + 40
5 A
0.0001 A
1.5 % + 20
0.8 % + 20
3 % + 40
[4]
Not Spec’d
AC A
[2]
10 A
[1]
0.001 A
1.5 % + 5
0.8 % + 5
3 % + 10
[4]
Not Spec’d
[1]
10 A range (10 % to 100 % of range).
[2]
20 A for 30 seconds on, 10 minutes off. >10 A not specified.
[3]
400 mA continuous; 550 mA for 2 minutes on, 1 minute off.
[4]
Verified by design and type tests.