Electrical (cont.) Measurement Accuracy, T1, T2, or T1-T2 (Model 52) Temperature Coefficient Electromagnetic Compatibility Maximum Differential Common Mode Voltage Temperature Scale J-, K-, T-, reading + 0a.n3doEC-t(y0p.5e:o±F[)0] .05 % of [below −100 oC (−148 oF): add 0.15 % of reading for J-, K-, E-, and N-type; and 0.45 % of reading for T-type] [ts0(obp.0e0e+.l10co85iw2%fieoo−FodF1f)p+0rre1e0ar8anoodCgFoienC)(g−ot1uo+4ts208i8.d0oeoF3C)to:hC(ae+d6pd4eroFoC 0.04 % of reading for J-, K-, E-, and N-type; and 0.08 % of reading for T-type] Susceptibility: ±2 oC (±3.6 oF) for 80 MHz to 200 MHz in 1.5 V/m field, for 200 MHz to 1000 MHz in 3 V/m field. Emmisions: Commercial Limits per EN50081-1 1 V (Maximum voltage difference between T1 and T2) ITS-90 Applicable Standards NIST-175 Accuracy is between 18 soCpe(c6i4fieodF)foarnadm2b8ieonCt (t8e2mopFe)rafotur raesperiod of 1 year. The above specifications do not include thermocouple error. 51 & 52 Series II Replacement Parts and Accessories Replacement Parts and Accessories Accessory Part Number Holster and Flex Stand™ Assembly 1272438 AA NEDA 15A IEC LR6 batteries 376756 80PK-1 K-Type Bead Thermocouple 773135 CD-ROM 1276106 Service Manual 1276123 15

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20

51 & 52 Series II
Replacement Parts and Accessories
15
Electrical (cont.)
Measurement
Accuracy, T1,
T2, or T1-T2
(Model 52)
J-, K-, T-, and E-type:
±
[0.05
%
of
reading
+
0.3
o
C (0.5
o
F)]
[below
-
100
o
C (
-
148
o
F): add 0.15 %
of reading for J-, K-, E-, and N-type;
and 0.45 % of reading for T-type]
Temperature
Coefficient
0.01
%
of reading
+
0.03
o
C per
o
C
(0.05
o
F per
o
F) outside the
specified
+
18
o
C to 28
o
C (
+
64
o
F
to
+
82
o
F) range
[below
-
100
o
C (
-
148
o
F): add
0.04 % of reading for J-, K-, E-, and
N-type; and 0.08 % of reading for
T-type]
Electromagnetic
Compatibility
Susceptibility:
±
2
o
C (
±
3.6
o
F) for 80
MHz to 200 MHz in 1.5 V/m field, for
200 MHz to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1
Maximum
Differential
Common Mode
Voltage
1 V (Maximum voltage difference
between T1 and T2)
Temperature
Scale
ITS-90
Applicable
Standards
NIST-175
Accuracy is specified for ambient temperatures
between 18
o
C (64
o
F) and 28
o
C (82
o
F) for a period of
1 year. The above specifications do not include
thermocouple error.
Replacement Parts and Accessories
Accessory
Part Number
Holster and Flex Stand
Assembly
1272438
AA NEDA 15A IEC LR6 batteries
376756
80PK-1 K-Type Bead Thermocouple
773135
CD-ROM
1276106
Service Manual
1276123