Operating the Thermometer - emissivity table
View all Fluke 561 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
Infrared Thermometer Operating the Thermometer 132 mm @ 1500 mm 75 mm @ S 900 mm 38 mm @ 300 mm 1.5 " @ 12 " 3 " @ 36 " 5.3 " @ 60 " D D:S=12:1 Figure 5. Distance and Spot Size efh005f.eps Field of View Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it. See Figure 6. Yes No Figure 6. Field of View efh004f.eps Emissivity Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or oxidized surfaces have an emissivity of about 0.95. If possible, to compensate for inaccurate readings that may result from measuring shiny metal surfaces, cover the surface to be measured with masking tape or flat black paint (< 148 °C/300 °F) and use the high emissivity setting. Allow time for the tape or paint to reach the same temperature as the surface beneath it. Measure the temperature of the tape or painted surface. If you cannot paint or use tape, then you can improve the accuracy of your measurements with the emissivity selector. Even with the emissivity selector, it can be difficult to get a completely accurate infrared measurement of a target with a shiny or metallic surface. Experimentation, using the probe to determine benchmark temperatures, and experience will help you choose the best setting for specific measurements. The Thermometer has three emissivity settings: low (0.3), medium (0.7), and high (0.95). Refer to Table 2. The reference to emissivity settings in the table are suggestions for typical situations. Your particular situation may differ. 7