Fluke 718-300G Fluke 717 and 718 Pressure Calibrator Datasheet - Page 2

Functional Specifications - calibration

Page 2 highlights

The New 718 Calibrator Pump The new Fluke 718 is a uniquely protected pneumatic pressure calibrator and pump. The new pump design uses protective check valves that allow air to pass through the pump for calibration but inhibits fluids from entering the pump and causing damage. These two check valves allow the bi-directional flow of air into or out of the 718. Even after draining the process fluid from the calibration port of the device under test, minute amounts of process fluids may often still be lurking behind. In the event these fluids enter the 718 pump cavities, the new pump design enables quick cleaning and maintenance, even in the field. Pump valve assembly cleaning instructions 1. Using a small screwdriver, remove the two valve retention caps located in the oval shaped opening on the underside of the Calibrator. 2. After the caps have been removed, gently remove the spring and o-ring assembly. 3. Set aside the valve assemblies in a safe area and clean out the valve body using a cotton swab soaked in IPA (isopropyl alcohol). 4. Repeat this process several times using a new cotton swab each time until there is no remaining sign of residue. 5. Pump the unit several times and check again for residue. 6. Clean the o-ring assembly and o-ring on the Functional Specifications retention caps with IPA and inspect the o-rings closely for any cuts, nicks, or wear. Replace if needed. 7. Inspect the springs for wear or loss of tension. They should be approximately 8.6 mm long in the relaxed state. If they are shorter than this, they may not allow the o-ring to seat properly. Replace if needed. 8. Once all parts have been cleaned and inspected, reinstall the o- ring and spring assemblies into the valve body. 9. Reinstall the retention caps Cleaning the fluid trap in the Fluke-718 and gently Calibrator is quick, easy and only requires a tighten the cap. slotted screwdriver and cotton swab 10. Seal the output of the Calibrator and pump up the unit to at least 50 % its rated pressure. 11. Release the pressure and repeat several times to ensure that the o-rings seat properly. The Calibrator is now ready for use. Functional Pressure Specifications Model Range 718 1G -1 PSI to +1 PSI, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa) 718 30G -12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa) 718 100G -12 PSI to 100 PSI, (-850 mbar to 6.895 bar, -85 to 689.48 kPa) 718 300G -12 PSI to 300 PSI, (-850 mbar to 20.68 bar, -85 to 2068.42 kPa ) 717 1G -1 PSI to 1 PSI, (-68.9 mbar to 68.9 mbar, -6.89 to 6.89 kPa) 717 30G -12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa) 717 100G -12 PSI to 100 PSI, (-850 mbar to 6.895 bar, -85 to 689.48 kPa) 717 300G -12 PSI to 300 PSI, (-850 mbar to 20.68 bar, -85 to 2068.4 kPa) 717 500G 0 PSI to 500 PSI, (0 mbar to 34.47 bar, 0 to 3447.4 kPa ) 717 1000G 0 PSI to 1000 PSI, (0 mbar to 68.95 bar, 0 to 6894.8 kPa ) 717 1500G 0 PSI to 1500 PSI, (0 mbar to 103.42 bar, 0 to 10342 kPa) 717 3000G 0 PSI to 3000 PSI, (0 mbar to 206.84 bar, 0 to 20684 kPa) 717 5000G 0 PSI to 5000 PSI, (0 mbar to 344.74 bar, 0 to 34474 kPa) Resolution 0.0001 psi, 0.001 mbar 0.001 psi, 0.1 mbar 0.01 psi, 1 mbar 0.01 psi, 1 mbar 0.001 psi, 0.001 mbar 0.001 psi, 0.1 mbar 0.01 psi, 1 mbar 0.01 psi, 1 mbar 0.01 psi, 1 mbar 0.1 psi, 1 mbar 0.1 psi, 0.01 bar 0.1 psi, 0.01 bar 0.1 psi, 0.01 bar 2 Fluke Pressure Calibrators Over Pressure Over Pressure 5xFS Over Pressure 2xFS Over Pressure 2xFS Over Pressure 375 PSI 25 bar Over Pressure 5xFS Over Pressure 2xFS Over Pressure 2xFS Over Pressure 375 PSI, 25 bar Over Pressure 2xFS Over Pressure 2xFS Over Pressure 2xFS Over Pressure 2xFS Over Pressure 2xFS Functions Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp Zero, Min, Max, Hold, Damp

  • 1
  • 2
  • 3

2
Fluke Pressure Calibrators
Functional Pressure Specifications
Model
Range
Resolution
Over Pressure
Functions
718 1G
-1 PSI to +1 PSI, -68.9 mbar to
68.9 mbar (-6.89 kPa to 6.89 kPa)
0.0001 psi, 0.001 mbar
Over Pressure 5xFS
Zero, Min, Max, Hold, Damp
718 30G
-12 PSI to 30 PSI, (-850 mbar to
2 bar, -85 to 206.84 kPa)
0.001 psi, 0.1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
718 100G
-12 PSI to 100 PSI, (-850 mbar to
6.895 bar, -85 to 689.48 kPa)
0.01 psi, 1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
718 300G
-12 PSI to 300 PSI, (-850 mbar to
20.68 bar, -85 to 2068.42 kPa )
0.01 psi, 1 mbar
Over Pressure 375 PSI
25 bar
Zero, Min, Max, Hold, Damp
717 1G
-1 PSI to 1 PSI, (-68.9 mbar to
68.9 mbar, -6.89 to 6.89 kPa)
0.001 psi, 0.001 mbar
Over Pressure 5xFS
Zero, Min, Max, Hold, Damp
717 30G
-12 PSI to 30 PSI, (-850 mbar to
2 bar, -85 to 206.84 kPa)
0.001 psi, 0.1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 100G
-12 PSI to 100 PSI, (-850 mbar to
6.895 bar, -85 to 689.48 kPa)
0.01 psi, 1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 300G
-12 PSI to 300 PSI,
(-850 mbar to
20.68 bar, -85 to 2068.4 kPa)
0.01 psi, 1 mbar
Over Pressure 375 PSI,
25 bar
Zero, Min, Max, Hold, Damp
717 500G
0 PSI to 500 PSI,
(0 mbar to
34.47 bar, 0 to 3447.4 kPa )
0.01 psi, 1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 1000G
0 PSI to 1000 PSI,
(0 mbar to
68.95 bar, 0 to 6894.8 kPa )
0.1 psi, 1 mbar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 1500G
0 PSI to 1500 PSI,
(0 mbar to
103.42 bar, 0 to 10342 kPa)
0.1 psi, 0.01 bar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 3000G
0 PSI to 3000 PSI,
(0 mbar to
206.84 bar, 0 to 20684 kPa)
0.1 psi, 0.01 bar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
717 5000G
0 PSI to 5000 PSI,
(0 mbar to
344.74 bar, 0 to 34474 kPa)
0.1 psi, 0.01 bar
Over Pressure 2xFS
Zero, Min, Max, Hold, Damp
Functional Specifications
The New 718 Calibrator Pump
The new Fluke 718 is a uniquely protected pneu-
matic pressure calibrator and pump. The new pump
design uses protective check valves that allow air to
pass through the pump for calibration but inhibits
fluids from entering the pump and causing damage.
These two check valves allow the bi-directional
flow of air into or out of the 718. Even after drain-
ing the process fluid from the calibration port of the
device under test, minute amounts of process fluids
may often still be lurking behind. In the event these
fluids enter the 718 pump cavities, the new pump
design enables quick cleaning and maintenance,
even in the field.
Pump valve assembly cleaning instructions
1. Using a small screwdriver, remove the two valve
retention caps located in the oval shaped opening
on the underside of the Calibrator.
2. After the caps have been removed, gently remove
the spring and o-ring assembly.
3. Set aside the valve assemblies in a safe area
and clean out the valve body using a cotton swab
soaked in IPA (isopropyl alcohol).
4. Repeat this process several times using a new
cotton swab each time until there is no remaining
sign of residue.
5. Pump the unit several times and check again for
residue.
6. Clean the o-ring assembly and o-ring on the
retention caps with IPA and inspect the o-rings
closely for any cuts, nicks, or wear. Replace if
needed.
7. Inspect the springs for wear or loss of tension.
They should be approximately 8.6 mm long in the
relaxed state. If
they are shorter
than this, they
may not allow
the o-ring to
seat properly.
Replace if
needed.
8. Once all
parts have
been cleaned
and inspected,
reinstall the o-
ring and spring
assemblies into
the valve body.
9. Reinstall the
retention caps
and gently
tighten the cap.
10. Seal the output of the Calibrator and pump up
the unit to at least 50 % its rated pressure.
11. Release the pressure and repeat several times to
ensure that the o-rings seat properly.
The Calibrator is now ready for use.
Cleaning the fluid trap in the Fluke-718
Calibrator is quick, easy and only requires a
slotted screwdriver and cotton swab