Using an External Pump, Table 4. Recommended Pressure Modules
View all Fluke 718Ex-100G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
Table 4. Recommended Pressure Modules Pressure Module 700P01Ex 700P24Ex 700P05Ex 700P06Ex 700P27Ex 700P09Ex 700PA4Ex 700P29Ex External Pump 718Ex 30G/100G X X X X X X X X Internal Pump 718Ex 30G 718Ex 100G X X X X X X X X X Pressure Calibrator Calibrating a P/I Transmitter Using an External Pump XW Warning To avoid damage to the Calibrator and possible release of pressure, do not connect the internal sensor to an external pressure source that exceeds 30 psi for Model 718Ex 30G or 100 psi for Model 718Ex 100G. To develop higher pressure or vacuum, use an external pump. Use a Fluke 700PEx Pressure Module connected to the pressure module input on the Calibrator. Pressure modules are listed in Table 4. Make overall connections as shown in Figure 6. Refer to setup and operating instructions included with the pressure module and pump. 17