971 Users Manual Specifications Temperature Range: Accuracy: Resolution: Update rate: Sensor type: Relative Humidity Range: Accuracy: Resolution: Response time: Sensor hysteresis: Sensor type: Temperature Coefficient: Wet Bulb Temperature Range: Dew Point Temperature Range: -20 to 60 °C (-4 to 140 °F) ±0.5 °C on 0 to 45 °C ±1.0 °C on -20 to 0 °C, 45 to 60 °C ±1.0 °F on 32 to 113 °F ±2.0 °F on -4 to 32 °F, 113 to 140 °F 0.1 °C /°F 500 ms NTC 5 to 95 % RH ±2.5 % RH (10 to 90 % RH) @23 °C (73.4 °F) ±5.0 % RH (90 % RH) @23 °C (73.4 °F) 0.1 % RH 60 seconds max. ±1 % RH with excursion of 90 % to 10 % to 90 % Electronic-capacitance polymer film 0.1 x (specified accuracy)/°C (< 23 °C or > 23 °C) -20 to 60 °C (-4 to 140 °F) -50 to 60 °C (-58 to 140 °F) 8

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12

971
Users Manual
8
Specifications
Temperature
Range:
Accuracy:
Resolution:
Update rate:
Sensor type:
-20 to 60 °C (-4 to 140 °F)
±0.5 °C on 0 to 45 °C
±1.0 °C on -20 to 0 °C, 45 to 60 °C
±1.0 °F on 32 to 113 °F
±2.0 °F on -4 to 32 °F, 113 to 140 °F
0.1 °C /°F
500 ms
NTC
Relative Humidity
Range:
Accuracy:
Resolution:
Response time:
Sensor hysteresis:
Sensor type:
Temperature
Coefficient:
5 to 95 % RH
±2.5 % RH (10 to 90 % RH) @23 °C
(73.4 °F)
±5.0 % RH (<10, >90 % RH) @23 °C
(73.4 °F)
0.1 % RH
60 seconds max.
±1 % RH with excursion of 90 % to 10 %
to 90 %
Electronic–capacitance polymer film
0.1 x (specified accuracy)/°C (< 23 °C or
> 23 °C)
Wet Bulb Temperature
Range:
-20 to 60 °C (-4 to 140 °F)
Dew Point Temperature
Range:
-50 to 60 °C (-58 to 140 °F)