Fluke BT521 Product Manual - Page 19
Accuracy Specifications, Function, Range, Resolution, Accuracy
View all Fluke BT521 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 19 highlights
BT521 Users Manual Accuracy Specifications Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F), with relative humidity to 80 %. Accuracy specifications are given as: ±([% of reading] + [number of least significant digits]). Accuracy specification assumes ambient temperature stable ±1 °C. Function Range Resolution Accuracy Battery Internal Resistance[1] 3 mΩ 30 mΩ 300 mΩ 3000 mΩ 6 V 0.001 mΩ 0.01 mΩ 0.1 mΩ 1 mΩ 0.001 V 1 % + 8 0.8 % + 6 0.8 % + 6 0.8 % + 6 V dc 60 V 600 V 0.01 V 0.1 V 0.09 % + 5 1000 V 1 V V ac (45 Hz to 500 Hz with lowpass filter) 600 V 0.1 V 2 % + 10 Frequency (Display with V ac, A ac using i410)[2] Trigger level: ≥ 10 mV @V ac; ≥ 10 A @A ac 45 Hz to 500 Hz 0.1 Hz 0.5 % + 8 AC Voltage Ripple (20 kHz max) 600 mV 6000 mV 0.1 mV 1 mV 3 % + 20 3 % + 10 Amps dc/Amps ac (With accessory Fluke i410) 400 A 1 A 3.5 % + 2 Temperature 0 °C to 60 °C 1 °C 2 °C (typical) [1] The measurement is based on ac injection method. The injected source signal is