Fluke BT521 Product Manual - Page 19

Accuracy Specifications, Function, Range, Resolution, Accuracy

Page 19 highlights

BT521 Users Manual Accuracy Specifications Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F), with relative humidity to 80 %. Accuracy specifications are given as: ±([% of reading] + [number of least significant digits]). Accuracy specification assumes ambient temperature stable ±1 °C. Function Range Resolution Accuracy Battery Internal Resistance[1] 3 mΩ 30 mΩ 300 mΩ 3000 mΩ 6 V 0.001 mΩ 0.01 mΩ 0.1 mΩ 1 mΩ 0.001 V 1 % + 8 0.8 % + 6 0.8 % + 6 0.8 % + 6 V dc 60 V 600 V 0.01 V 0.1 V 0.09 % + 5 1000 V 1 V V ac (45 Hz to 500 Hz with lowpass filter) 600 V 0.1 V 2 % + 10 Frequency (Display with V ac, A ac using i410)[2] Trigger level: ≥ 10 mV @V ac; ≥ 10 A @A ac 45 Hz to 500 Hz 0.1 Hz 0.5 % + 8 AC Voltage Ripple (20 kHz max) 600 mV 6000 mV 0.1 mV 1 mV 3 % + 20 3 % + 10 Amps dc/Amps ac (With accessory Fluke i410) 400 A 1 A 3.5 % + 2 Temperature 0 °C to 60 °C 1 °C 2 °C (typical) [1] The measurement is based on ac injection method. The injected source signal is

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66

BT521
Users Manual
1-12
Accuracy Specifications
Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F
to 82 °F), with relative humidity to 80 %. Accuracy specifications are given as: ±([% of
reading] + [number of least significant digits]). Accuracy specification assumes ambient
temperature stable ±1 °C.
Function
Range
Resolution
Accuracy
Battery Internal Resistance
[1]
3 m
0.001 m
1 % + 8
30 m
0.01 m
0.8 % + 6
300 m
0.1 m
0.8 % + 6
3000 m
1 m
0.8 % + 6
V dc
6 V
0.001 V
0.09 % + 5
60 V
0.01 V
600 V
0.1 V
1000 V
1 V
V ac (45 Hz to 500 Hz with low-
pass filter)
600 V
0.1 V
2 % + 10
Frequency (Display with V ac,
A ac using i410)
[2]
Trigger level:
10 mV @V ac;
10 A @A ac
45 Hz to 500 Hz
0.1 Hz
0.5 % + 8
AC Voltage Ripple (20 kHz
max)
600 mV
6000 mV
0.1 mV
1 mV
3 % + 20
3 % + 10
Amps dc/Amps ac (With
accessory Fluke i410)
400 A
1 A
3.5 % + 2
Temperature
0 °C to 60 °C
1 °C
2 °C (typical)
[1]
The measurement is based on ac injection method. The injected source signal is <100 mA, 1 kHz.
[2]
For A ac frequency measurement, the range is 45 Hz to 400 Hz.