Fluke Calibration 1620A-H-156 Getting Started Guide - Page 11

Specifications and Environmental Conditions, Specifications

Page 11 highlights

2 Specifications and Environmental Conditions Specifications and Environmental Conditions Specifications Temperature Range Temperature Accuracy ("H" Model) Temperature Accuracy ("S" Model) Delta Temperature Accuracy Temperature Display Resolution RH Range RH Accuracy ("H" Model) RH Accuracy ("S" Model) Delta Humidity Accuracy RH Display Resolution Inputs Display Memory Alarms Alarm port output Communications Ethernet Wireless Range Enclosure Power Battery Backup Operating Range Size (DewK)HxWxD Size (Probes) Weight 0 °C to 50 °C 16 °C to 24 °C (60.8 °F to 75.2 °F): ± 0.125 °C (± 0.225 °F) [calibrated] 0 °C to 16 °C (32 °F to 60.8 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical] 24 °C to 50 °C (75.2 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical] 15 °C to 35 °C (59 °F to 95 °F): ± 0.25 °C (± 0.45 °F) [calibrated] 0 °C to 15 °C (32 °F to 59 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical] 35 °C to 50 °C (95 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical] ± 0.025 °C (± 0.045 °F) for ± 1 °C (± 1.8 °F) changes within 15 °C to 35 °C (59 °F to 95 °F) User selectable up to 0.001 °C (0.01 °C recorded) 0 % to 100 %RH 20 % to 70 %RH: ± 1.5 %RH (calibrated) 0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical) 20 % to 70 %RH: ± 2 %RH (calibrated) 0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical) ± 1.0% for ± 5% changes within 20% to 70 %RH User selectable up to 0.01% (0.1% recorded) Two sensors, each measuring temperature and relative humidity; each is detachable, cable-extendable, and interchangeable, with self-contained calibration; each may be assigned a unique 16-character identification 240 x 128 graphics monochrome LCD, displays temperature and humidity data graphically, numerically, and statistically; 16 pre-defined, userchangeable screen set-ups are included 400,000 typical individual time-stamped readings Visual and audio alarms for temperature, temperature rate, RH, RH rate, and fault conditions 0 V normal, 11 to 12 V active, sources up to 20mA, 2.5mm, two-conductor subminiature plug RS-232, Ethernet LAN, 802.15.4 (ZigBee) wireless (optional) 10 Base-T, 100 Base-TX, IP, TCP, DHCP, Ping, HTTP, HTML 30 m (100 ft) typical unobstructed The DewK may be wall-mounted (hardware included) or set on a benchtop 12 V DC from external 100-240 V AC power supply Standard 9V battery to allow continued measuring during power disruptions 0 °C to 50 °C 125 mm x 211 mm x 51 mm (4.9 in x 8.3 in x 2.0 in) 79 mm H x 19 mm dia. (3.1in x 0.75 in) 0.7 kg (1.5 lb.) 

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26

7
2 Specifications and Environmental Conditions
Specifications and Environmental Conditions
Specifications
Temperature Range
0 °C to 50 °C
Temperature Accuracy
(“H” Model)
16 °C to 24 °C (60.8 °F to 75.2 °F): ± 0.125 °C (± 0.225 °F) [calibrated]
0 °C to 16 °C (32 °F to 60.8 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
24 °C to 50 °C (75.2 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
Temperature Accuracy
(“S” Model)
15 °C to 35 °C (59 °F to 95 °F): ± 0.25 °C (± 0.45 °F) [calibrated]
0 °C to 15 °C (32 °F to 59 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
35 °C to 50 °C (95 °F to 122 °F): ± 0.5 °C (± 0.9 °F) [uncalibrated typical]
Delta Temperature
Accuracy
± 0.025 °C (± 0.045 °F) for ± 1 °C (± 1.8 °F)
changes within 15 °C to 35 °C (59 °F to 95 °F)
Temperature Display
Resolution
User selectable up to 0.001 °C (0.01 °C recorded)
RH Range
0 % to 100 %RH
RH Accuracy
(“H” Model)
20 % to 70 %RH: ± 1.5 %RH (calibrated)
0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical)
RH Accuracy
(“S” Model)
20 % to 70 %RH: ± 2 %RH (calibrated)
0 % to 20 %RH, 70% to 100 %RH: ± 3 %RH (uncalibrated, typical)
Delta Humidity
Accuracy
± 1.0% for ± 5% changes within 20% to 70 %RH
RH Display Resolution
User selectable up to 0.01% (0.1% recorded)
Inputs
Two sensors, each measuring temperature and relative humidity; each is
detachable, cable-extendable, and interchangeable, with self-contained
calibration; each may be assigned a unique 16-character identification
Display
240 x 128 graphics monochrome LCD, displays temperature and humidity
data graphically, numerically, and statistically; 16 pre-defined, user-
changeable screen set-ups are included
Memory
400,000 typical individual time-stamped readings
Alarms
Visual and audio alarms for temperature, temperature rate, RH, RH rate,
and fault conditions
Alarm port output
0 V normal, 11 to 12 V active, sources up to 20mA, 2.5mm, two-conductor
subminiature plug
Communications
RS-232, Ethernet LAN, 802.15.4 (ZigBee) wireless (optional)
Ethernet
10 Base-T, 100 Base-TX, IP, TCP, DHCP, Ping, HTTP, HTML
Wireless Range
30 m (100 ft) typical unobstructed
Enclosure
The DewK may be wall-mounted (hardware included) or set on a benchtop
Power
12 V DC from external 100-240 V AC power supply
Battery Backup
Standard 9V battery to allow continued measuring during power disruptions
Operating Range
0 °C to 50 °C
Size (DewK)HxWxD
125 mm x 211 mm x 51 mm (4.9 in x 8.3 in x 2.0 in)
Size (Probes)
79 mm H x 19 mm
dia. (3.1in x 0.75 in)
Weight
0.7 kg (1.5 lb.)