4 Usage 4.1 Before Use WARNING! Risk of Injury or Damage Defective or inoperable cylinders may cause injury or damage. - All cylinders MUST be inspected before attempting to fill. - If the cylinder is found to be defective or does not work properly, contact an Invacare dealer or qualified technician for service or replacement. Before filling a cylinder assembly, read and follow the information in Cylinder Prefill Inspection and External Examination. 4.2 Connecting/Disconnecting the Cylinder Assembly to/from the Compressor Refer to the HomeFill Compressor user manual for operating instructions to connect and disconnect the cylinder to the HomeFill Compressor. 4.3 Cylinder Assembly Operation DANGER! Risk of Fire - NEVER smoke in an area where oxygen is being administered. - NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. 1112648-H-00 Usage 1. Prior to each use, inspect product for visible damage. Refer to the External Examination section. DO NOT use if any damage is found. If for any reason, any label becomes illegible or lost, contact your dealer/Home Care equipment provider. 2. Inspect the cylinder gauge F to ensure the pressure is at least 1500 psig (10342 kPa). If below 1500 psig (10342 kPa), then refill the cylinder assembly using the HomeFill system. WARNING! - When cylinder gauge needle enters red zone, it is recommended to refill the cylinder assembly on the Invacare HomeFill Compressor. 17