Invacare® ML4, ML6 and M9 Cylinder with Conserver 8 Technical Data 8.1 Technical Description The Invacare Oxygen Conserving Cylinder is an oxygen storage product that can deliver up to 5 L/min of pulse flow oxygen to the patient via a nasal cannula. The cylinder assembly is designed to store compressed oxygen up to 2,100 psig (14479 kPa). The conserver reduces the pressure at the cannula connection to 50 psig (345 kPa) and contains a rotary flow dial that is used to adjust the oxygen flow rate to the prescribed setting. The cylinder assembly is filled with USP 93±3% oxygen using its proprietary fill port that allows filling only through the use of an Invacare HomeFill System. The duration of oxygen flow between fills is dependent on the size of the cylinder and the flow setting. 8.2 Specifications Settings Conserver S/N Range: Pulse Settings: Flow Settings: Savings Ratio: 1-69,999 1, 2, 3, 4, & 5 2 L/min Continuous Up to 4.3:1 75,000+ 1, 2, 3, 4, & 5 2 L/min Continuous Up to 5.7:1 Pressures Conserver S/N Range: Inlet Pressure Range: Internal Regulated Pressure: Max Fill Pressure: Burst Pressure: 1-69,999 300-2100 psig 50 psig 2100 psig (14479 kPa) 3025-3360 psig at 72°F (20857-23166 kPa at 22.2°C) 75,000+ 300-2100 psig 50 psig 2100 psig (14479 kPa) 3025-3360 psig at 72°F (20857-23166 kPa at 22.2°C) 24 1112648-H-00