Invacare® ML4, ML6 and M9 Cylinder with Conserver CAUTION! Risk of Damage - DO NOT completely empty the cylinder. Otherwise, damage to the cylinder could occur. 1. Position the cylinder assembly A so the outlet connection D is pointing away from the user and any other person(s). 2. Turn the flow dial to setting "Off". 3. Turn the flow dial E to setting "5". 4. Examine the cylinder gauge F until the pressure is below 500 psig (3447 kPa). Cylinder pressure is below 500 psig (3447 kPa) when the cylinder gauge needle is in the red zone. 5. Turn the flow dial E to setting "Off". 20 5.3 Cleaning Refer to the safety information in the Maintenance section before performing this procedure. Follow the cleaning/disinfectant product manufacturer's instructions for use, handling, storage and disposal of the product. Follow this procedure to clean the cylinder assembly as required when visibly soiled. 1. Ensure the cover is installed on the fillport. 2. Clean exterior surfaces of the cylinder with a lint free cloth dampened with 70% isopropyl alcohol. 3. Clean the exterior surfaces of the conserver with a dry, lint free cloth. 4. If the conserver is still visibly soiled, clean with a lint free cloth dampened with 70% isopropyl alcohol. DO NOT wipe the fillport or outlet fitting. 5. Wipe the cylinder and conserver with a dry lint free cloth. Ensure the product is dry before using. 5.4 Disinfecting Refer to the safety information in the Maintenance section before performing this procedure. Follow the cleaning/disinfectant product manufacturer's instructions for use, handling, storage and disposal of the product. This procedure is for disinfecting the cylinder assembly and is to be performed by the provider prior to a patient change and as otherwise required. Recommended disinfectant products: • Quaternary ammonium based wipes selected by the provider for the desired disinfecting characteristics 1112648-H-00

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28

Invacare® ML4, ML6 and M9 Cylinder with Conserver
CAUTION!
Risk of Damage
DO NOT completely empty the cylinder. Otherwise,
damage to the cylinder could occur.
1.
Position the cylinder assembly
A
so the outlet connection
D
is
pointing away from the user and any other person(s).
2.
Turn the flow dial to setting “Off”.
3.
Turn the flow dial
E
to setting “5”.
4.
Examine the cylinder gauge
F
until the pressure is below 500
psig (3447 kPa).
Cylinder pressure is below 500 psig (3447 kPa) when
the cylinder gauge needle is in the red zone.
5.
Turn the flow dial
E
to setting “Off”.
5.3
Cleaning
Refer to the safety information in the Maintenance section
before performing this procedure.
Follow the cleaning/disinfectant product manufacturer’s instructions
for use, handling, storage and disposal of the product.
Follow this procedure to clean the cylinder assembly as required
when visibly soiled.
1.
Ensure the cover is installed on the fillport.
2.
Clean exterior surfaces of the cylinder with a lint free cloth
dampened with 70% isopropyl alcohol.
3.
Clean the exterior surfaces of the conserver with a dry, lint
free cloth.
4.
If the conserver is still visibly soiled, clean with a lint free cloth
dampened with 70% isopropyl alcohol. DO NOT wipe the
fillport or outlet fitting.
5.
Wipe the cylinder and conserver with a dry lint free cloth.
Ensure the product is dry before using.
5.4
Disinfecting
Refer to the safety information in the Maintenance section
before performing this procedure.
Follow the cleaning/disinfectant product manufacturer’s instructions
for use, handling, storage and disposal of the product.
This procedure is for disinfecting the cylinder assembly and is to be
performed by the provider prior to a patient change and as otherwise
required.
Recommended disinfectant products:
Quaternary ammonium based wipes selected by the provider
for the desired disinfecting characteristics
20
1112648-H-00