Invacare IRC5PO2V Owners Manual - Page 12

Typical Product, Parameters - oxygen concentrator

Page 12 highlights

SECTION 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS Electrical Requirements: Rated Current Input: Sound Level: Altitude: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz 3 A 43 ± 2 dBA Average for Perfecto2 Models and Perfecto2V Model 40 ± 2 dBA Average for Perfecto2W Models All Perfecto2 Models: Up to 8,000 ft (2438 meters) above sea level without degredation of concentration levels. From 8,000 ft to 13,129 (4000 meters) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency. 87% to 95.6% for flows ranging at 0.5 to 5 L/min 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa) 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Perfecto2 and Perfecto2W Models: 280 W (Operating @ 3 L/min) Pressure Relief Mechanism Operational at: Perfecto2V Model: 325 W 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) Perfecto2™Series 12 Part No 1143482

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

SECTION 4—TYPICAL PRODUCT PARAMETERS
Perfecto
2
Series
12
Part No 1143482
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
3 A
Sound Level:
43 ± 2 dBA Average for Perfecto
2
Models and
Perfecto
2
V Model
40 ± 2 dBA Average for Perfecto
2
W Models
Altitude:
All Perfecto
2
Models:
Up to 8,000 ft (2438
meters) above sea level without degredation of
concentration levels. From 8,000 ft to 13,129
(4000 meters) below 90% efficiency.
All Perfecto
2
Models while filling a
HomeFill System:
Up to 6,000 ft (1828 meters)
above sea level without degradation of
concentration levels. From 6,000 ft (1828 meters)
to 13,129 ft (4000 meters) below 90% efficiency.
*Oxygen Output
Concentration Levels:
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pres-
sure:
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range:
0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Low Flow Alarm:
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories
Connected).
Average Power
Consumption:
Perfecto
2
and Perfecto
2
W Models: 280 W
(Operating @ 3 L/min)
Perfecto
2
V Model: 325 W
Pressure Relief Mecha-
nism Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)