Invacare IRC5PO2V Owners Manual - Page 26

Troubleshooting, Guide

Page 26 highlights

SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep...... PROBABLE CAUSE: Main Power Loss: 1. Power cord not plugged in. 2. No power at outlet. 3. Tripped circuit breaker. SOLUTION: 1. Insert plug into outlet. 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet. 3. Call Service Provider. Alarm: Continuous Concentrator not operating, Power Switch On. System Failure: 1. Unit overheating due to blocked air intake. Beeeeeeeep.... 2. Insufficient power at outlet. 3. Internal repairs required. 1a. Remove and clean cabinet filters. 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. DO NOT use extension cords. Move to another electrical outlet or circuit. 3. Call Service Provider. Alarm: Rapid Beep....Beep... Beep...Beep Low Flow Alarm: 1. Kinked or blocked tubing, cannula or humidifier. 2. Flowmeter set at 0.5 L/min or less. 1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back ON. 2. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Perfecto2™Series 26 Part No 1143482

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

SECTION 7—TROUBLESHOOTING GUIDE
Perfecto
2
Series
26
Part No 1143482
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Short beeps, long
pause
Concentrator not
operating,
power switch On.
Beep
......
Beep
......
Main Power Loss:
1. Power cord not
plugged in.
1. Insert plug into outlet.
2. No power at outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
3. Tripped circuit
breaker.
3. Call Service Provider.
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep
....
System Failure:
1. Unit overheating due
to blocked air intake.
1a. Remove and clean cabinet
filters.
1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.
2. Insufficient power at
outlet.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Internal repairs
required.
3. Call Service Provider.
Alarm:
Rapid
Beep
....
Beep...
Beep...Beep
Low Flow Alarm:
1. Kinked or blocked
tubing, cannula or
humidifier.
1. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
2. Flowmeter set at 0.5
L/min or less.
2. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only
with Perfecto
2
5).