Invacare IRC5PO2V Owners Manual - Page 26
Troubleshooting, Guide
View all Invacare IRC5PO2V manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 26 highlights
SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep...... PROBABLE CAUSE: Main Power Loss: 1. Power cord not plugged in. 2. No power at outlet. 3. Tripped circuit breaker. SOLUTION: 1. Insert plug into outlet. 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet. 3. Call Service Provider. Alarm: Continuous Concentrator not operating, Power Switch On. System Failure: 1. Unit overheating due to blocked air intake. Beeeeeeeep.... 2. Insufficient power at outlet. 3. Internal repairs required. 1a. Remove and clean cabinet filters. 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. DO NOT use extension cords. Move to another electrical outlet or circuit. 3. Call Service Provider. Alarm: Rapid Beep....Beep... Beep...Beep Low Flow Alarm: 1. Kinked or blocked tubing, cannula or humidifier. 2. Flowmeter set at 0.5 L/min or less. 1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back ON. 2. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Perfecto2™Series 26 Part No 1143482