Invacare M91-TS Owners Manual 3 - Page 55
Front View of Footplate
View all Invacare M91-TS manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 55 highlights
Angle Adjustment For this procedure, refer to FIGURE 4 on page 54 and FIGURE 5. Observe the angle of the footplate for reinstallation. The settings for positioning the footplates on the half-clamps may vary for each footplate. 1. Loosen, but DO NOT remove, the two flat screws, washer and locknuts that secure the footplate to the footplate hinge (FIGURE 4). 2. Position the footplate to the necessary angle to accommodate the user (FIGURE 5). 3. Retighten the two flat screws, washers and locknuts. Perpendicular and/or Inversion/Eversion Adjustment For this procedure, refer to FIGURE 4 on page 54 and FIGURE 6. It is not necessary to remove the footplate to perform this adjustment. 1. Insert a flathead screwdriver through the half clamp on the footplate (FIGURE 4). 2. Slowly turn nylon adjustment screw in or out until footplate is perpendicular to the footrest assembly or the desired inversion or eversion is obtained (FIGURE 6). Footrest Support Footplate 6 ADJUSTMENT Side View of Footplate and Footrest Support FIGURE 5 Angle Adjustment Front View of Footplate and Footrest Support Footrest Support Footplate FIGURE 6 Perpendicular and/or Inversion/Eversion Adjustment Part No 1143192 55 Invacare® Adjustable ASBA Seats