Sealey MM20 Instruction Manual - Page 3

Specifications, Maintenance

Page 3 highlights

5. specifications DC VOLTAGE Range Accuracy Resolution 200mV 2V 20V 200V 600V ±0.5% of reading ±2 digits ±0.8% of reading ±3 digits ±0.8% of reading ±3 digits ±0.8% of reading ±3 digits ±0.8% of reading ±5 digits 100µV 1mV 10mV 100mV 1V Overload protection: 220V rms AC for 200mV range and 600V DC or 600V rms AC for other ranges. AC VOLTAGE Range 200V 600V Accuracy ± 2% of rdg ± 2 digits ± 2% of rdg ± 2 digits Resolution 100mV 1V Overload protection: 600V DC or 600V rms AC. Response: Average responding, calibrated in rms of a sine wave. Frequency range: 40Hz - 400Hz RESISTANCE Range Accuracy Resolution 200Ω ±0.8% of reading ±5 digits 0.1Ω 2kΩ ±0.8% of reading ±2 1Ω digits 20kΩ ±0.8% of reading ±2 10Ω digits 200kΩ ±0.8% of reading ±2 digits 100Ω 2MΩ ±1.2% of reading ±3 1kΩ digits Maximum Open Circuit Voltage: 2.8V Overload protection: 220V rms for a maximum of 10 seconds for all ranges. TEMPERATURE Range -40°C to 400°C Accuracy ± 1.0% of rdg ± 3°C 400°C to 1000°C ± 2.0% of rdg ± 3°C Overload protection: 220V rms AC Resolution 1°C 1°C DC CURRENT Range 2mA 20mA 200mA 10A Accuracy ±1.0% of reading ±2 digits ±1.0% of reading ±2 digits ±1.2% of reading ±2 digits ±2.0% of reading ±2 digits Overload protection: 200mA/250V fuse Measuring Voltage drop: 200mV Resolution 1µA 10µA 100µA 10mA TRANSISTOR hFE TEST Range (NPN or PNP) Displaying range 0 - 1000 Test Condition Current: 10µA Vce: 3V DIODE TEST Range Description Display Approx. Voltage Drop CONTINUITY TEST Range Description Buzzes if resistance is below 70Ω 5.1. ACCURACY CACULATION Example: Test reading on 200Vdc range is 56.4V. Accuracy is ±0.8% of reading ±3 digits. Reading ±3 digit = 56.4 ±3 on the last figure i.e. 56.1 to 56.7V. ±0.8% on this range gives 56.1 - 0.8% to 56.7 + 0.8% or 55.6 to 57.1V. Therefore the actual voltage lies between 55.6 and 57.1V. Note: Accuracy is specified for a period of one year after calibration and at 23°C±5°C with a relative humidity of less than 75%. 6. maintenancei ‰ ‰ 6.1. ‰ ‰ WARNING! Before attempting to open the case, ensure that the test leads have been disconnected from measurement circuits to avoid electric shock hazard. A fuse usually only blows due to operator error. To replace the fuse, remove the protective rubber boot and the two screws from the rear of the meter. Lift off the rear cover, replace the fuse and re-assemble the tester in reverse order, ensuring that the rear cover engages with the two lugs at the top of the tester. WARNING! ALWAYS replace a fuse with one of the correct rating (200mA/250V). © Jack Sealey Limited Original Language Version MM20.V3 | Issue:5 (I) 20/10/17

  • 1
  • 2
  • 3
  • 4

5. SPECIFICATIONS
5.1.
ACCURACY CACULATION
Example: Test reading on 200Vdc range is 56.4V. Accuracy is ±0.8% of reading ±3 digits. Reading ±3 digit = 56.4 ±3 on the last figure
i.e. 56.1 to 56.7V. ±0.8% on this range gives 56.1 - 0.8% to 56.7 + 0.8% or 55.6 to 57.1V.Therefore the actual voltage lies between 55.6
and 57.1V.
Note:
Accuracy is specified for a period of one year after calibration and at 23
°
C±5
°
C with a relative humidity of less than 75%.
6. MAINTENANCE
i
±
WARNING!
Before attempting to open the case, ensure that the test leads have been disconnected from measurement circuits to avoid
electric shock hazard.
6.1.
A fuse usually only blows due to operator error. To replace the fuse, remove the protective rubber boot and the two screws from the rear
of the meter. Lift off the rear cover, replace the fuse and re-assemble the tester in reverse order, ensuring that the rear cover engages
with the two lugs at the top of the tester.
±
WARNING! ALWAYS
replace a fuse with one of the correct rating (200mA/250V).
AC VOLTAGE
Range
Accuracy
Resolution
200V
± 2% of rdg ± 2 digits
100mV
600V
± 2% of rdg ± 2 digits
1V
Overload protection: 600V DC or 600V rms AC.
Response: Average responding, calibrated in rms of a
sine wave. Frequency range: 40Hz - 400Hz
TEMPERATURE
Range
Accuracy
Resolution
-40°C to 400°C
± 1.0% of rdg
± 3°C
1°C
400°C to 1000°C
± 2.0% of rdg
± 3°C
1°C
Overload protection: 220V rms AC
DIODE TEST
Range
Description
Display Approx.
Voltage Drop
TRANSISTOR
h
FE
TEST
Range
Displaying range
Test
Condition
(NPN or
PNP)
0 - 1000
Current: 10
µ
A
Vce: 3V
DC VOLTAGE
Range
Accuracy
Resolution
200mV
±0.5% of reading ±2 digits
100µV
2V
±0.8% of reading ±3 digits
1mV
20V
±0.8% of reading ±3 digits
10mV
200V
±0.8% of reading ±3 digits
100mV
600V
±0.8% of reading ±5 digits
1V
Overload protection: 220V rms AC for 200mV range
and 600V DC or 600V rms AC for other ranges.
DC CURRENT
Range
Accuracy
Resolution
2mA
±1.0% of reading ±2 digits
1µA
20mA
±1.0% of reading ±2 digits
10µA
200mA
±1.2% of reading ±2 digits
100
µ
A
10A
±2.0% of reading ±2 digits
10mA
Overload protection: 200mA/250V fuse
Measuring Voltage drop: 200mV
RESISTANCE
Range
Accuracy
Resolution
200
±0.8% of reading ±5
digits
0.1
2k
±0.8% of reading ±2
digits
1
20k
±0.8% of reading ±2
digits
10
200k
±0.8% of reading ±2
digits
100
2M
±1.2% of reading ±3
digits
1k
Maximum Open Circuit Voltage: 2.8V
Overload protection: 220V rms for a maximum of 10
seconds for all ranges.
CONTINUITY TEST
Range
Description
Buzzes if resistance
is below 70
Ω
MM20.V3 | Issue:5 (I) 20/10/17
Original Language Version
© Jack Sealey Limited